BIOMECHANICAL COMPARISON OF THE SLIDING HIP SCREW AND THE DOME PLUNGER

被引:25
作者
CHOUEKA, J [1 ]
KOVAL, KJ [1 ]
KUMMER, FJ [1 ]
CRAWFORD, G [1 ]
ZUCKERMAN, JD [1 ]
机构
[1] HOSP JOINT DIS & MED CTR, INST ORTHOPAED, DEPT BIOENGN, NEW YORK, NY 10003 USA
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1995年 / 77B卷 / 02期
关键词
D O I
10.1302/0301-620X.77B2.7706347
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We studied the biomechanical behaviour of three sliding fixation devices for trochanteric femoral fractures, These mere a titanium alloy sideplate and lag screw, a titanium alloy sideplate and dome plunger with cement augmentation, and a stainless-steel sideplate and lag screw. We used 18 mildly osteoporotic cadaver femora, randomly assigned to one of the three fixation groups, Four displacement and two strain gauges were fixed to each specimen, and each femur was first tested intact (control), then as a two-part fracture and then as a four-part intertrochanteric fracture. A range of physiological loads was applied to determine load-bearing, load-sharing and head displacement. The four-part-fracture specimens were subsequently tested to failure to determine maximum fixation strengths and modes of failure. The dome-plunger group failed at a load 50% higher than that of the stainless-steel lag-screw group (p < 0.05) and at a load 20% higher than that of the titanium-alloy lag-screw group (NS), All 12 lag-screw specimens failed by cut-out through the femoral head or neck, but none of the dome-plunger group showed movement within the femoral head when tested to failure, Strain-gauge analysis showed that the dome plunger produced considerably less strain in the inferior neck and calcar region than either of the lag screws, Inferior displacement of the femoral head was greatest for the dome-plunger group, and was due to sliding of the plunger. The dome plunger with cement augmentation was able to support higher loads and did not fail by cut-out through the femoral head, Its sliding potential was maintained, retaining the biomechanical advantages of the sliding screw.
引用
收藏
页码:277 / 283
页数:7
相关论文
共 30 条
[1]   THE EFFECT OF ADJUNCTIVE METHYLMETHACRYLATE ON FAILURES OF FIXATION AND FUNCTION IN PATIENTS WITH INTERTROCHANTERIC FRACTURES AND OSTEOPOROSIS [J].
BARTUCCI, EJ ;
GONZALEZ, MH ;
COOPERMAN, DR ;
FREEDBERG, HI ;
BARMADA, R ;
LAROS, GS .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1985, 67A (07) :1094-1107
[2]   DENSITY OF FRESH AND EMBALMED HUMAN COMPACT AND CANCELLOUS BONE [J].
BLANTON, PL ;
BIGGS, NL .
AMERICAN JOURNAL OF PHYSICAL ANTHROPOLOGY, 1968, 29 (01) :39-&
[3]  
CHANG WS, 1987, CLIN ORTHOP RELAT R, V225, P141
[4]   LONG-TERM RESULTS AND COMPLICATIONS OF CEMENT AUGMENTATION IN THE TREATMENT OF UNSTABLE TROCHANTERIC FRACTURES [J].
CHENG, CL ;
CHOW, SP ;
PUN, WK ;
LEONG, JCY .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1989, 20 (03) :134-138
[5]   FEMORAL-NECK FRACTURE FIXATION - COMPARISON OF A SLIDING SCREW WITH LAG SCREWS [J].
CLARK, DI ;
CROFTS, CE ;
SALEH, M .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (05) :797-800
[6]   INTERTROCHANTERIC FEMORAL FRACTURES - MECHANICAL FAILURE AFTER INTERNAL-FIXATION [J].
DAVIS, TRC ;
SHER, JL ;
HORSMAN, A ;
SIMPSON, M ;
PORTER, BB ;
CHECKETTS, RG .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (01) :26-31
[7]   TREATMENT OF THE UNSTABLE INTERTROCHANTERIC FRACTURE - EFFECT OF THE PLACEMENT OF THE SCREW, ITS ANGLE OF INSERTION, AND OSTEOTOMY [J].
DENHARTOG, BD ;
BARTAL, E ;
COOKE, F .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1991, 73A (05) :726-733
[8]   UNSTABLE INTERTROCHANTERIC FRACTURES OF HIP [J].
DIMON, JH ;
HUGHSTON, JC .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1967, A 49 (03) :440-&
[9]  
DOHERTY JH, 1979, CLIN ORTHOP RELAT R, V141, P184
[10]  
EVANS FG, 1948, ANAT REC, V100, P159