DIAGNOSIS OF EARLY CARDIAC TRANSPLANT REJECTION BY FALL IN EVOKED T-WAVE AMPLITUDE MEASURED USING AN EXTERNALIZED QT DRIVEN RATE RESPONSIVE PACEMAKER

被引:18
作者
GRACE, AA
NEWELL, SA
CARY, NRB
SCOTT, JP
LARGE, SR
WALLWORK, J
SCHOFIELD, PM
机构
[1] PAPWORTH HOSP,REG CARDIOL UNIT,CAMBRIDGE CB3 8RE,ENGLAND
[2] PAPWORTH HOSP,TRANSPLANT UNIT,CAMBRIDGE CB3 8RE,ENGLAND
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1991年 / 14卷 / 06期
关键词
CARDIAC TRANSPLANTATION; EVOKED RESPONSE; RATE RESPONSIVE PACEMAKER;
D O I
10.1111/j.1540-8159.1991.tb04153.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Reliable diagnosis of cardiac allograft rejection is at present only possible using endomyocardial biopsy. We have serially measured epicardial evoked T wave amplitude during ventricular pacing with an externalized QT driven rate responsive pacemaker telemetered to a TP2 analyzer in 13 patients (12 males) followed for 19 (14-26) days after transplantation. A total of 228 records were analyzed. Rejection was defined on endomyocardial biopsy. On 17 of the 31 occasions on which biopsy was performed during the study, specimens showed significant (moderate) rejection. In 11 patients the initial biopsy proven rejection episode was associated with a significant fall in the evoked T wave amplitude from 1.3 (0.7-2.3) mV to 0.6 (0.5-1.8) mV (P < 0.005), which began 2 (1-4) days earlier. One patient with uncontrolled diabetes mellitus had no change in evoked T wave amplitude during rejection. The evoked T wave amplitude did not fall in the absence of histologic rejection. These results suggest a noninvasive method for detecting cardiac rejection, which appears both sensitive (92%) and specific (100%) in the first rejection episodes.
引用
收藏
页码:1024 / 1031
页数:8
相关论文
共 12 条
[1]  
BAIG MW, 1989, BRIT HEART J, V61, P473
[2]  
Billingham ME., 1982, HEART TRANSPLANTATIO, V1, P25
[3]   RECENT EXPERIENCE WITH HEART-TRANSPLANTATION [J].
ENGLISH, TAH ;
COOPER, DKC ;
CORYPEARCE, R .
BMJ-BRITISH MEDICAL JOURNAL, 1980, 281 (6242) :699-702
[4]   RELIABILITY OF THE EVOKED-RESPONSE IN DETERMINING THE PACED VENTRICULAR RATE AND PERFORMANCE OF THE QT OR RATE RESPONSIVE (TX) PACEMAKER [J].
FANANAPAZIR, L ;
RADEMAKER, M ;
BENNETT, DH .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1985, 8 (05) :701-714
[5]   FREQUENCY-ANALYSIS OF THE SURFACE ELECTROCARDIOGRAM FOR RECOGNITION OF ACUTE REJECTION AFTER ORTHOTOPIC CARDIAC TRANSPLANTATION IN MAN [J].
HABERL, R ;
WEBER, M ;
REICHENSPURNER, H ;
KEMKES, BM ;
OSTERHOLZER, G ;
ANTHUBER, M ;
STEINBECK, G .
CIRCULATION, 1987, 76 (01) :101-108
[6]  
JAMIESON SW, 1979, BRIT HEART J, V42, P703
[7]  
KEMKES BM, 1989, HEART HEART LUNG TRA, P299
[8]  
KEREN A, 1984, CIRCULATION, V70, P124
[9]   ELECTROCARDIOGRAMS OF DOGS WITH HEART HOMOGRAFTS [J].
LOWER, RR ;
DONG, E ;
GLAZENER, FS .
CIRCULATION, 1966, 33 (03) :455-&
[10]  
RICKARDS AF, 1981, BRIT HEART J, V45, P56