SINGLE-CHAMBER VERSUS DUAL-CHAMBER SENSOR-DRIVEN PACING - COMPARISON OF CARDIAC OUTPUTS

被引:12
作者
PROCTOR, EE
LEMAN, RB
MANN, DL
KAISER, J
KRATZ, J
GILLETTE, P
机构
[1] MED UNIV S CAROLINA,DEPT MED,DIV CARDIOL,171 ASHLEY AVE,CHARLESTON,SC 29425
[2] MED UNIV S CAROLINA,DEPT SURG,DIV CARDIOTHORAC SURG,CHARLESTON,SC 29425
[3] MED UNIV S CAROLINA,DEPT PEDIAT,DIV PEDIAT CARDIOL,CHARLESTON,SC 29425
[4] VET AFFAIRS MED CTR,CHARLESTON,SC 29425
关键词
D O I
10.1016/0002-8703(91)90518-M
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Previous studies have shown that single-chamber sensor-driven pacing improves exercise tolerance for patients with chronotropic incompetence. However, long-term single-chamber pacing has a number of inherent problems that limit its usefulness. Although sensor-driven dual-chamber pacing largely obviates the problems inherent with single-chamber sensor-driven pacing, the physiologic benefit of dual-chamber sensor-driven pacing has not yet been demonstrated. Accordingly, the purpose of this study was to compare exercise-induced cardiac output for patients with chronotropic incompetence, after programming their pacemakers to either a simulated sensor-driven single or simulated dual-chamber mode. Cardiac output was measured noninvasively at rest and peak exercise using standard Doppler-derived measurements, obtained in a blinded fashion. At rest the Doppler-derived resting VVI and DDD cardiac outputs were 4.49 +/- 0.3 L/min and 4.68 +/- 0.3 L/min, respectively. At peak exercise, the DDD cardiac output was 5.07 +/- 0.5 L/min, whereas the simulated activity VVI and DDD cardiac outputs were 6.33 +/- 0.6 L/min and 7.41 +/- 0.70 L/min, respectively. Analysis of variance showed that there was an overall significant difference in cardiac output from rest to peak exercise (p < 0.001). However, only the simulated activity DDD cardiac output was significantly different from its respective control value (p < 0.05). Thus this study shows for the first time that the addition of rate responsiveness to dual-chamber pacing results in a significant improvement in cardiac output for patients with chronotropic incompetence.
引用
收藏
页码:728 / 732
页数:5
相关论文
共 15 条
[1]  
AUSUBEL K, 1985, PACE, V8, P318
[2]   SINGLE-CHAMBER CARDIAC PACING WITH ACTIVITY-INITIATED CHRONOTROPIC RESPONSE - EVALUATION BY CARDIOPULMONARY EXERCISE TESTING [J].
BENDITT, DG ;
MIANULLI, M ;
FETTER, J ;
BENSON, DW ;
DUNNIGAN, A ;
MOLINA, E ;
GORNICK, CC ;
ALMQUIST, A .
CIRCULATION, 1987, 75 (01) :184-191
[3]   DECREASE IN CANINE ENDOCARDIAL AND EPICARDIAL ELECTROGRAM VOLTAGES WITH EXERCISE - IMPLICATIONS FOR PACEMAKER SENSING [J].
BRICKER, JT ;
WARD, KA ;
ZINNER, A ;
GILLETTE, PC .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (04) :460-464
[4]   EFFECT OF VENTRICULAR-FUNCTION ON THE EXERCISE HEMODYNAMICS OF VARIABLE-RATE PACING [J].
BUCKINGHAM, TA ;
WOODRUFF, RC ;
PENNINGTON, DG ;
REDD, RM ;
JANOSIK, DL ;
LABOVITZ, AJ ;
GRAVES, R ;
KENNEDY, HL .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1988, 11 (06) :1269-1277
[5]  
FAERESTRAND S, 1987, PACE, V10, P673
[6]   USEFULNESS OF TRANS-CUTANEOUS TRIGGERING OF CONVENTIONAL IMPLANTED PULSE GENERATORS BY AN ACTIVITY-SENSING PACEMAKER FOR PREDICTING EFFECTIVENESS OF RATE RESPONSE PACING [J].
FETTER, J ;
BENDITT, DG ;
MIANULLI, M .
AMERICAN JOURNAL OF CARDIOLOGY, 1988, 62 (13) :901-905
[7]   ATRIAL SIGNAL VARIATIONS AND PACEMAKER MALSENSING DURING EXERCISE - A STUDY IN THE TIME AND FREQUENCY-DOMAIN [J].
FROHLIG, G ;
SCHWERDT, H ;
SCHIEFFER, H ;
BETTE, L .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1988, 11 (04) :806-813
[8]   EVALUATION OF PULMONARY AND SYSTEMIC BLOOD-FLOW BY TWO-DIMENSIONAL DOPPLER ECHOCARDIOGRAPHY USING FAST FOURIER-TRANSFORM SPECTRAL-ANALYSIS [J].
GOLDBERG, SJ ;
SAHN, DJ ;
ALLEN, HD ;
VALDESCRUZ, LM ;
HOENECKE, H ;
CARNAHAN, Y .
AMERICAN JOURNAL OF CARDIOLOGY, 1982, 50 (06) :1394-1400
[9]  
GWINN N, 1990, PACE, V13, P535
[10]   RATE RESPONSIVE DUAL CHAMBER PACING [J].
KAPPENBERGER, LJ ;
HERPERS, L .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1986, 9 (06) :987-991