ACCURACY OF PATIENT POSITIONING IN MANTLE FIELD IRRADIATION

被引:19
作者
CREUTZBERG, CL [1 ]
VISSER, AG [1 ]
DEPORRE, PMZR [1 ]
MEERWALDT, JH [1 ]
ALTHOF, VGM [1 ]
LEVENDAG, PC [1 ]
机构
[1] DR DANIEL DEN HOED CANC CTR,DEPT CLIN PHYS,3075 EA ROTTERDAM,NETHERLANDS
关键词
MANTLE FIELD IRRADIATION; LOCALIZATION ERRORS; PATIENT POSITIONING; QUALITY CONTROL;
D O I
10.1016/S0167-8140(92)80131-2
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
A prospective study of the accuracy of patient positioning in mantle field irradiation was carried out in 13 lymphoma patients treated with curative radiotherapy. Patients were treated in the supine and prone position for anterior and posterior fields, respectively. Individually shaped divergent shielding blocks were placed in a fixed position in a template which was positioned on a tray above the patient. A total number of 94 megavoltage portal films (MV) was analysed and compared to 26 simulation films (SIM). MV-SIM differences were larger for posterior fields than for anterior fields. Regarding the position of the lung shielding blocks, mean MV-SIM differences ranged from 1.3 to 4.4 mm and errors exceeding 1 cm were found in 7.2% of cases. Most discrepancies appeared to be randomly distributed. A 4-5 mm systematic cranial shift of patients in the posterior treatment position was noted. Discrepancies in the position of the laryngeal block, spinal cord shielding block and humerus blocks were small with mean MV-SIM differences ranging from 0.3 to 2.7 mm. Differences between simulation set-up and treatment set-up were modest as compared to error rates reported in the literature. Shielding of tumour-bearing areas did not occur. It was concluded that the present standardised technique of patient positioning and the design of treatment fields results in acceptable error rates. Attention should be directed towards increasing the stability of patients in the prone treatment position in order to further reduce both systematic and random error rates.
引用
收藏
页码:257 / 264
页数:8
相关论文
共 16 条
  • [1] WEEKLY LOCALIZATION FILMS AND DETECTION OF FIELD PLACEMENT ERRORS
    BYHARDT, RW
    COX, JD
    HORNBURG, A
    LIERMANN, G
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1978, 4 (9-10): : 881 - 887
  • [2] GLATSTEIN E, 1987, PRINCIPLES PRACTICE, P1057
  • [3] Grant L, 1973, Clin Radiol, V24, P254, DOI 10.1016/S0009-9260(73)80094-7
  • [4] THE DAILY REPRODUCIBILITY OF LARGE, COMPLEX-SHAPED RADIOTHERAPY FIELDS TO THE THORAX AND NECK
    GRIFFITHS, SE
    PEARCEY, RG
    [J]. CLINICAL RADIOLOGY, 1986, 37 (01) : 39 - 41
  • [5] ACCURACY IN RADIATION-FIELD ALIGNMENT IN HEAD AND NECK-CANCER - A PROSPECTIVE-STUDY
    HUIZENGA, H
    LEVENDAG, PC
    DEPORRE, PMZR
    VISSER, AG
    [J]. RADIOTHERAPY AND ONCOLOGY, 1988, 11 (02) : 181 - 187
  • [6] LOCALIZATION ERRORS IN MANTLE-FIELD IRRADIATION FOR HODGKINS-DISEASE
    HULSHOF, M
    VANUYTSEL, L
    VANDENBOGAERT, W
    VANDERSCHUEREN, E
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1989, 17 (03): : 679 - 683
  • [7] KINZIE JJ, 1983, CANCER, V52, P2223, DOI 10.1002/1097-0142(19831215)52:12<2223::AID-CNCR2820521209>3.0.CO
  • [8] 2-D
  • [9] MARKS JE, 1976, CANCER, V37, P2755, DOI 10.1002/1097-0142(197606)37:6<2755::AID-CNCR2820370628>3.0.CO
  • [10] 2-J