FLOW-INJECTION TECHNIQUE FOR THE DETERMINATION OF LOW-LEVELS OF PHOSPHORUS IN NATURAL-WATERS

被引:39
作者
FREEMAN, PR [1 ]
MCKELVIE, ID [1 ]
HART, BT [1 ]
CARDWELL, TJ [1 ]
机构
[1] LA TROBE UNIV,DEPT CHEM,ANALYT CHEM LABS,BUNDOORA,VIC 3083,AUSTRALIA
基金
澳大利亚研究理事会;
关键词
Phosphorus; Waters;
D O I
10.1016/S0003-2670(00)83584-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An inexpensive flow-injection instrument for determining low concentrations of dissolved reactive phosphorus in natural waters is reported. The unique feature is the use of an inexpensive detector consisting of a flow cell and a simple photometer that incorporates a super-bright light-emitting diode as the source and a photodiode as the detector. The tin(II) chloride-molybdate method was optimized using a modified simplex optimization method. Silicate interference up to 5 mg Si l-1 was removed by addition of 0.10% (w/v) tartaric acid. Using the tartaric acid-modified optimized reagents, a detection limit of 0.6 μg P l-1 was achieved. The method was linear over the range 0-100 μg P l-1 with an excellent precision (r.s.d. 2.9% at 2.0 and 0.5% at 50 μg P l-1). An in-line pre-concentration anion-exchange column was used to obtain an even lower detection limit of 0.1 μg P l-1 and applied to the analysis of real samples. © 1990.
引用
收藏
页码:409 / 416
页数:8
相关论文
共 23 条