FIELD-STRENGTH IN NEURO-MR IMAGING - A COMPARISON OF 0.5-T AND 1.5-T

被引:28
作者
JACK, CR [1 ]
BERQUIST, TH [1 ]
MILLER, GM [1 ]
FORBES, GS [1 ]
GRAY, JE [1 ]
MORIN, RL [1 ]
ILSTRUP, DM [1 ]
机构
[1] MAYO CLIN & MAYO FDN,DEPT HLTH SCI RES,ROCHESTER,MN 55905
关键词
Brain; Comparative studies; Magnetic resonance imaging; Spine; Techniques;
D O I
10.1097/00004728-199007000-00002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A study was undertaken comparing neurological magnetic resonance imaging at high (1.5 T) and mid (0.5 T) field strengths. Twenty-eight patients (20 head and 8 spine) from our routine case load volunteered to undergo two consecutive and identical MR studies on the two systems. The two MR systems were built by the same manufacturer and were equipped with essentially identical hardware and software. Individual patient studies were performed consecutively in adjacent MR suites, and pulse sequence parameters were replicated exactly at the two field strengths. One exception to this rule was that the second echo of the long TR sequence in the head was acquired with a narrow receiver bandwidth on the 0.5 T system. The resulting axial double echo long repetition time (TR) and sagittal short TR head images and sagittal short and double echo long TR spine images were graded by two blinded observers (senior staff neuroradiologists) on two levels. First, the images were graded for image quality, i.e., conspicuousness of artifacts and clarity in depiction of normal and pathologic anatomy. Second, diagnostic accuracy of MR was assessed relative to the clinical-pathologic diagnosis in each case. The image quality of the 1.5 T system was rated superior in both the head and spine for most specific items assessed. This observer preference for 1.5 T images did not, however, translate into greater diagnostic accuracy for the 1.5 T system in the head. Although the 1.5 T system did have a slight advantage in diagnostic accuracy in the spine, a significant difference was not found. © 1988 Raven Press, Ltd., New York.
引用
收藏
页码:505 / 513
页数:9
相关论文
共 17 条
  • [1] CEREBRAL MAGNETIC-RESONANCE - COMPARISON OF HIGH AND LOW FIELD-STRENGTH IMAGING
    BILANIUK, LT
    ZIMMERMAN, RA
    WEHRLI, FW
    GOLDBERG, HI
    GROSSMAN, RI
    BOTTOMLEY, PA
    EDELSTEIN, WA
    GLOVER, GH
    MACFALL, JR
    REDINGTON, RW
    KRESSEL, HY
    [J]. RADIOLOGY, 1984, 153 (02) : 409 - 414
  • [2] MAGNETIC-RESONANCE IMAGING OF PITUITARY LESIONS USING 1.0 TO 1.5-T FIELD-STRENGTH
    BILANIUK, LT
    ZIMMERMAN, RA
    WEHRLI, FW
    SNYDER, PJ
    GOLDBERG, HI
    GROSSMAN, RI
    BOTTOMLEY, PA
    EDELSTEIN, WA
    GLOVER, GH
    MACFALL, JR
    REDINGTON, RW
    [J]. RADIOLOGY, 1984, 153 (02) : 415 - 418
  • [3] RF MAGNETIC-FIELD PENETRATION, PHASE-SHIFT AND POWER DISSIPATION IN BIOLOGICAL TISSUE - IMPLICATIONS FOR NMR IMAGING
    BOTTOMLEY, PA
    ANDREW, ER
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1978, 23 (04) : 630 - 643
  • [4] CENTRAL NERVOUS-SYSTEM HIGH-RESOLUTION MAGNETIC-RESONANCE IMAGING - EFFECT OF INCREASING SPATIAL-RESOLUTION ON RESOLVING POWER
    BRADLEY, WG
    KORTMAN, KE
    CRUES, JV
    [J]. RADIOLOGY, 1985, 156 (01) : 93 - 98
  • [5] THE FIELD-DEPENDENCE OF NMR IMAGING .1. LABORATORY ASSESSMENT OF SIGNAL-TO-NOISE RATIO AND POWER DEPOSITION
    CHEN, CN
    SANK, VJ
    COHEN, SM
    HOULT, DI
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (05) : 722 - 729
  • [6] NUCLEAR MAGNETIC-RESONANCE WHOLE-BODY IMAGER OPERATING AT 3.5 KGAUSS
    CROOKS, L
    ARAKAWA, M
    HOENNINGER, J
    WATTS, J
    MCREE, R
    KAUFMAN, L
    DAVIS, PL
    MARGULIS, AR
    DEGROOT, J
    [J]. RADIOLOGY, 1982, 143 (01) : 169 - 174
  • [7] MAGNETIC-RESONANCE IMAGING - EFFECTS OF MAGNETIC-FIELD STRENGTH
    CROOKS, LE
    ARAKAWA, M
    HOENNINGER, J
    MCCARTEN, B
    WATTS, J
    KAUFMAN, L
    [J]. RADIOLOGY, 1984, 151 (01) : 127 - 133
  • [8] DRAYER B, 1986, AM J NEURORADIOL, V7, P373
  • [9] NMR RELAXATION-TIMES OF BLOOD - DEPENDENCE ON FIELD-STRENGTH, OXIDATION-STATE, AND CELL INTEGRITY
    GOMORI, JM
    GROSSMAN, RI
    YUIP, C
    ASAKURA, T
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1987, 11 (04) : 684 - 690
  • [10] NUCLEAR MAGNETIC-RESONANCE IMAGING - CONTRAST-TO-NOISE RATIO AS A FUNCTION OF STRENGTH OF MAGNETIC-FIELD
    HART, HR
    BOTTOMLEY, PA
    EDELSTEIN, WA
    KARR, SG
    LEUE, WM
    MUELLER, O
    REDINGTON, RW
    SCHENCK, JF
    SMITH, LS
    VATIS, D
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1983, 141 (06) : 1195 - 1201