COMPARISON OF A PULSED ELECTRON-CAPTURE DETECTOR AND A FOURIER-TRANSFORM ION MOBILITY DETECTOR AFTER CAPILLARY SUPERCRITICAL FLUID CHROMATOGRAPHY

被引:9
作者
TARVER, EE [1 ]
HILL, HH [1 ]
机构
[1] WASHINGTON STATE UNIV,DEPT CHEM,PULLMAN,WA 99164
来源
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY | 1992年 / 344卷 / 10-11期
关键词
D O I
10.1007/BF00323742
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A commercial electron capture detector and an ion mobility detector were interfaced to a supercritical fluid chromatograph (SFC). The ion mobility detector (IMD) was operated in the negative ion collection mode wherein analyte molecules of high electron affinity form negatively charged ions in the detector, producing an ECD-like response. Responses of the two detectors were compared using pressure programmed SFC with carbon dioxide as the mobile phase. A mixture of nine chlorinated pesticides (Lindane, DDT, DDE, BHC, Endrin, Heptachlor, Endosulfan, Dieldrin and Chloropicrin); a mixture of six polycyclic aromatic hydrocarbons (naphthalene, acenaphthalene, dimethylnaphthalene, pyrene, phenanthrene, and fluoranthene); two polychlorinated biphenyls (Aroclor 1248 and Aroclor 1260), and a bean extract containing the chlorinated fungicide Captan were chromatographed as test compounds. In all cases pressure programming of the carbon dioxide mobile phase resulted in an increasing baseline signal as pressure increased with the ECD while a stable baseline was observed with the IMD. These investigations demonstrated the potential of negative mode IMD as a useful alternative for the sensitive and selective SFC detection of compounds amenable to electron capture detection.
引用
收藏
页码:453 / 459
页数:7
相关论文
共 31 条
[1]   TUNABLE SELECTIVE DETECTION FOR CAPILLARY GAS-CHROMATOGRAPHY BY ION MOBILITY MONITORING [J].
BAIM, MA ;
HILL, HH .
ANALYTICAL CHEMISTRY, 1982, 54 (01) :38-43
[2]   NEGATIVE IONS OF N2O AND CO2 [J].
BARDSLEY, JN .
JOURNAL OF CHEMICAL PHYSICS, 1969, 51 (08) :3384-&
[3]   COMPARISON OF THE NEGATIVE REACTANT IONS FORMED IN THE PLASMA CHROMATOGRAPH BY NITROGEN, AIR, AND SULFUR-HEXAFLUORIDE AS THE DRIFT GAS WITH AIR AS THE CARRIER GAS [J].
CARR, TW .
ANALYTICAL CHEMISTRY, 1979, 51 (06) :705-711
[4]  
CHANG HCK, 1990, J CHROMATOGR SCI, V28, P29
[5]  
DRESSLER M, 1986, J CHROMATOGR LIBRARY, V36
[6]   IRON-55 AS AN AUGER-ELECTRON EMITTER - NOVEL SOURCE FOR GAS-CHROMATOGRAPHY DETECTORS [J].
DWIGHT, DJ ;
LORCH, EA ;
LOVELOCK, JE .
JOURNAL OF CHROMATOGRAPHY, 1976, 116 (02) :257-261
[7]   CAPILLARY SUPERCRITICAL FLUID CHROMATOGRAPHY WITH CONVENTIONAL FLAME DETECTORS [J].
FJELDSTED, JC ;
KONG, RC ;
LEE, ML .
JOURNAL OF CHROMATOGRAPHY, 1983, 279 (NOV) :449-455
[8]   CHARACTERIZATION OF MICROWAVE-INDUCED PLASMA AS A DETECTOR FOR SUPERCRITICAL FLUID CHROMATOGRAPHY [J].
GALANTE, LJ ;
SELBY, M ;
LUFFER, DR ;
HIEFTJE, GM ;
NOVOTNY, M .
ANALYTICAL CHEMISTRY, 1988, 60 (14) :1370-1376
[10]   ALTERATION OF THE ELECTRON-CAPTURE DETECTOR RESPONSE TO POLYCYCLIC AROMATIC-HYDROCARBONS BY OXYGEN DOPING OF THE CARRIER GAS [J].
GRIMSRUD, EP ;
MILLER, DA ;
STEBBINS, RG ;
KIM, SH .
JOURNAL OF CHROMATOGRAPHY, 1980, 197 (01) :51-58