AUDIBILITY AND IDENTIFICATION OF AUDITORY ALARMS IN THE OPERATING-ROOM AND INTENSIVE-CARE UNIT

被引:78
作者
MOMTAHAN, K [1 ]
HETU, R [1 ]
TANSLEY, B [1 ]
机构
[1] UNIV MONTREAL,ECOLE ORTHOPHONIE & AUDIOL,ACOUST GRP,MONTREAL H3C 3J7,QUEBEC,CANADA
关键词
MONITORING; PHYSIOLOGICAL; EQUIPMENT DESIGN; STANDARDS; INTENSIVE CARE; OPERATING ROOMS; ANESTHESIA;
D O I
10.1080/00140139308967986
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The audibility and the identification of 23 auditory alarms in the intensive care unit (ICU) and 26 auditory alarms in the operating rooms (ORs) of a 214-bed Canadian teaching hospital were investigated. Digital tape recordings of the alarms were made and analysed using masked-threshold software developed at the Universite de Montreal. The digital recordings were also presented to the hospital personnel responsible for monitoring these alarms on an individual basis in order to determine how many of the alarms they would be able to identify when they heard them. Several of the alarms in both areas of the hospital could mask other alarms in the same area, and many of the alarms in the operating rooms could be masked by the sound of a surgical saw or a surgical drill. The staff in the OR (anaesthetists, anaesthesia residents, and OR technologists) were able to identify a mean of between 10 and 15 of the 26 alarms found in their operating theatres. The ICU nurses were able to identify a mean of between 9 and 14 of the 23 alarms found in their ICU. Alarm importance was positively correlated with the frequency of alarm identification in the case of the OR, rho = 0.411, but was not significantly correlated in the case of the ICU, rho = 0.155. This study demonstrates the poor design of auditory warning signals in hospitals and the need for standardization of alarms on medical equipment.
引用
收藏
页码:1159 / 1176
页数:18
相关论文
共 43 条
[1]  
*AM SOC TEST MAT, 1991, SPEC AL SIGN MED EQ
[2]  
*ASS ADV MED INSTR, 1988, HUM FACT ENG GUID PR, P8
[3]   EFFECTS OF AUDITORY PITCH AND COMPLEXITY ON EEG DESYNCHRONIZATION AND ON VERBALLY EXPRESSED JUDGEMENTS [J].
BERLYNE, DE ;
MCDONNEL. ;
NICKI, RM ;
PARHAM, LCC .
CANADIAN JOURNAL OF PSYCHOLOGY, 1967, 21 (04) :346-346
[4]   LISTENER EVALUATION OF SIMULATED TELEPHONE CALLING SIGNALS [J].
BRICKER, PD .
BELL SYSTEM TECHNICAL JOURNAL, 1971, 50 (05) :1559-+
[5]  
De Boer E., 1976, HDB SENSORY PHYSL, P479
[6]  
Deatherage Bruce H., 1972, HUMAN ENG GUIDE EQUI, P123
[7]   AUDITORY SIGNALS IN MILITARY AIRCRAFT - ERGONOMICS PRINCIPLES VERSUS PRACTICE [J].
DOLL, TJ ;
FOLDS, DJ .
APPLIED ERGONOMICS, 1986, 17 (04) :257-264
[8]   IMPROVING AUDITORY WARNING DESIGN - RELATIONSHIP BETWEEN WARNING SOUND PARAMETERS AND PERCEIVED URGENCY [J].
EDWORTHY, J ;
LOXLEY, S ;
DENNIS, I .
HUMAN FACTORS, 1991, 33 (02) :205-231
[9]   PERCEIVED URGENCY AND THE ANESTHETIST - RESPONSES TO COMMON OPERATING-ROOM MONITOR ALARMS [J].
FINLEY, GA ;
COHEN, AJ .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1991, 38 (08) :958-964