DETERMINATION OF TRACE ANIONS IN ISOPROPANOL

被引:13
作者
KAISER, E
WOJTUSIK, MJ
机构
[1] Dionex Corporation, Sunnyvale, CA 94088
关键词
D O I
10.1016/0021-9673(94)80247-5
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Ion chromatography along with matrix elimination reliably determines low mu g/l levels of anionic contaminants in isopropanol. The use of ion-exchange columns that are solvent compatible makes elimination of the isopropanol sample matrix possible and allows calibration standards to be prepared in deionized water. By concentrating a large amount of sample, detection limits for chloride, sulfate, phosphate and nitrate between 0.2 and 1.0 mu g/l are achieved. The sample can be injected without sample pretreatment. The procedure is automated and yields a complete analysis in less than 1 h.
引用
收藏
页码:253 / 258
页数:6
相关论文
共 6 条
[1]   THE EVOLUTION OF SILICON-WAFER CLEANING TECHNOLOGY [J].
KERN, W .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1990, 137 (06) :1887-1892
[2]  
Poliak R., 1992, Microcontamination, V10, P45
[3]  
RABIN S, 1990, J CHROMATOGR, V640, P105
[4]  
RUNYAN WR, 1990, SEMICONDUCTOR INTEGR, P34
[5]  
STILLIAN JR, 1990, J CHROMATOGR, V499, P257
[6]  
1993, BOOK SEMI STANDARDS, P74