FRACTURE OF THE TIBIAL TRAY FOLLOWING PCA KNEE REPLACEMENT - A REPORT OF 2 CASES

被引:9
作者
MARUYAMA, M
TERAYAMA, K
SUNOHARA, H
ADACHI, T
SUZUKI, S
FUKUZAWA, T
机构
[1] Department of Orthopaedic Surgery, Shinshu University School of Medicine, Matsumoto, 390
关键词
D O I
10.1007/BF00426182
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We report two cases in which fracture of the tibial metal tray was thought to be due to improper design of the tray and heavy polyethylene wear. The bone beneath the fractured portion of the tray was deficient and had been replaced by granulomatous fibrous tissue including numerous polyethylene wear particles and foreign-body giant cells. Osteolysis occurred as a tissue reaction to these particles.
引用
收藏
页码:330 / 333
页数:4
相关论文
共 10 条
[1]
DANNENMAIER WC, 1985, CLIN ORTHOP RELAT R, V198, P224
[2]
FRACTURE OF THE TIBIAL TRAY OF THE PCA KNEE - A CASE-REPORT OF EARLY FAILURE CAUSED BY IMPROPER DESIGN [J].
FLIVIK, G ;
LJUNG, P ;
RYDHOLM, U .
ACTA ORTHOPAEDICA SCANDINAVICA, 1990, 61 (01) :26-28
[3]
Gradisar I A Jr, 1989, J Arthroplasty, V4, P27, DOI 10.1016/S0883-5403(89)80050-6
[5]
MENDES DG, 1984, ORTHOPEDICS, V7, P860, DOI 10.3928/0147-7447-19840501-12
[6]
AGGRESSIVE GRANULOMATOSIS FROM POLYETHYLENE FAILURE IN AN UNCEMENTED KNEE REPLACEMENT [J].
NOLAN, JF ;
BUCKNILL, TM .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (01) :23-24
[7]
PILLIAR RM, 1983, CLIN ORTHOP RELAT R, P42
[8]
RANAWAT CS, 1986, CLIN ORTHOP RELAT R, P244
[9]
FRACTURE OF THE METALLIC TIBIAL TRAY FOLLOWING TOTAL KNEE REPLACEMENT - REPORT OF 2 CASES [J].
SCOTT, RD ;
EWALD, FC ;
WALKER, PS .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1984, 66A (05) :780-782
[10]
SKINNER HB, 1987, ORTHOPEDICS, V10, P581