PERCUTANEOUS PEDESTAL IN COCHLEAR IMPLANTATION

被引:24
作者
PARKIN, JL
机构
[1] Division of Otolaryngology-Head and Neck Surgery, University of Utah School of Medicine, Salt Lake City, Utah
关键词
cochlear implant; percutaneous pedestal;
D O I
10.1177/000348949009901008
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Direct electrical connection between an external sound-processing device and intracochlear electrodes is accomplished via a percutaneous pedestal. Patient pedestal complaints and experience have been recorded. Pedestal problems have been classified into six classes, with class 0 indicating no problems and the severity of problem increasing in classes 1 through 4. Class 5 indicates pedestal trauma. No class 5 problems were encountered. At 48 months of experience, 53% of patients were class 0 and 26% were class 1. The pedestal experience is detailed in the paper. The percutaneous pedestal has been found to be a well-tolerated, efficient system for information transfer in cochlear implant patients. © 1990, SAGE Publications. All rights reserved.
引用
收藏
页码:796 / 800
页数:5
相关论文
共 4 条
[1]  
Holgers K M, 1987, Int J Oral Maxillofac Implants, V2, P35
[2]  
HOLGERS KM, 1988, AM J OTOL, V9, P56
[3]   SOFT-TISSUE INFECTION AROUND A SKIN PENETRATING OSSEOINTEGRATED IMPLANT [J].
JACOBSSON, M ;
TJELLSTROM, A ;
THOMSEN, P ;
ALBREKTSSON, T .
SCANDINAVIAN JOURNAL OF PLASTIC AND RECONSTRUCTIVE SURGERY AND HAND SURGERY, 1987, 21 (02) :225-228
[4]   INTEGRATION OF TITANIUM IMPLANTS IN IRRADIATED BONE - HISTOLOGIC AND CLINICAL-STUDY [J].
JACOBSSON, M ;
THOMSEN, P ;
ALBREKTSSON, T ;
TURESSON, I .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1988, 97 (04) :337-340