Bone loss around femoral hip stems is one of the problems threatening the long-term fixation of uncemented stems. Many believe that this phenomenon is caused by reduced stresses in the bone (stress shielding). In the present study the mechanical consequences of different femoral stem materials were investigated using adaptive bone remodeling theory in combination with the finite element method. Bone-remodeling in the femur around the implant and interface stresses between bone and implant were investigated for fully bonded femoral stems. Cemented stems (cobalt-chrome or titanium alloy) caused less bone resorption and lower interface stresses than uncemented stems made from the same materials. The range of the bone resorption predicted in the simulation models was from 23% in the proximal medial cortex surrounding the cemented titanium alloy stem to 76% in the proximal medial cortex around the uncemented cobalt-chrome stem. Very little bone resorption was predicted around a flexible, uncemented "iso-elastic" stem, but the proximal interface stresses increased drastically relative to the stiffer uncemented stems composed of cobalt-chrome or titanium alloy. However, the proximal interface stress peak was reduced and shifted during the adaptive remodeling process. The latter was found particularly in the stiffer uncemented cobalt-chrome-molybdenum implant and less for the flexible isoelastic implant.
引用
收藏
页码:845 / 853
页数:9
相关论文
共 31 条
[1]  
AHNFELT L, 1990, ACTA ORTHOP SCAND S, P238
[2]   AN APPROACH FOR TIME-DEPENDENT BONE MODELING AND REMODELING - APPLICATION - A PRELIMINARY REMODELING SIMULATION [J].
BEAUPRE, GS ;
ORR, TE ;
CARTER, DR .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1990, 8 (05) :662-670
[3]   RELATIONSHIPS BETWEEN LOADING HISTORY AND FEMORAL CANCELLOUS BONE ARCHITECTURE [J].
CARTER, DR ;
ORR, TE ;
FYHRIE, DP .
JOURNAL OF BIOMECHANICS, 1989, 22 (03) :231-244
[4]   COMPRESSIVE BEHAVIOR OF BONE AS A 2-PHASE POROUS STRUCTURE [J].
CARTER, DR ;
HAYES, WC .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1977, 59 (07) :954-962
[5]   BONE REMODELING .1. THEORY OF ADAPTIVE ELASTICITY [J].
COWIN, SC ;
HEGEDUS, DH .
JOURNAL OF ELASTICITY, 1976, 6 (03) :313-326
[6]  
ENGH CA, 1988, CLIN ORTHOPAEDICS, V23, P7
[7]  
Frost, 1964, LAWS BONE STRUCTURE
[9]   A UNIFYING PRINCIPLE RELATING STRESS TO TRABECULAR BONE MORPHOLOGY [J].
FYHRIE, DP ;
CARTER, DR .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1986, 4 (03) :304-317
[10]   ADAPTIVE BONE-REMODELING THEORY APPLIED TO PROSTHETIC-DESIGN ANALYSIS [J].
HUISKES, R ;
WEINANS, H ;
GROOTENBOER, HJ ;
DALSTRA, M ;
FUDALA, B ;
SLOOFF, TJ .
JOURNAL OF BIOMECHANICS, 1987, 20 (11-12) :1135-1150