ANTI-D QUANTIFICATION BY FLOW-CYTOMETRY - AN ALTERNATIVE TO THE AUTOANALYZER

被引:23
作者
AUSTIN, EB
MCINTOSH, Y
HODSON, C
LEE, D
机构
[1] Manchester Blood Centre, Manchester, M13 9LL, Plymouth Grove
关键词
ANTI-D; AUTOANALYZER; FLOW CYTOMETRY; QUANTIFICATION;
D O I
10.1111/j.1365-3148.1995.tb00229.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A how cytometry method for quantifying levels of anti-D is described. The method is based on the indirect antiglobulin test and uses an FITC Fab anti-human IgG reagent for detection of bound anti-D. A standard curve is generated from the fluorescence measured by flow cytometry using the British Standard anti-D preparation 73/515. The fluorescence obtained from test samples by flow cytometry is converted into IU/mL anti-D, using the standard curve. In a limited study with 42 serum samples the assay showed good correlation with results obtained using the AutoAnalyser for sera with anti-D levels less than 100 IU/mL. The method is simple to perform and has potential as a replacement for the AutoAnalyser.