CINEANGIOGRAPHY OF THE HEART IN A SINGLE BREATH HOLD WITH A SEGMENTED TURBOFLASH SEQUENCE

被引:289
作者
ATKINSON, DJ
EDELMAN, RR
机构
[1] BETH ISRAEL HOSP,DEPT RADIOL,330 BROOKLINE AVE,BOSTON,MA 02215
[2] SIEMENS MED SYST INC,ISELIN,NJ
关键词
HEART; MR STUDIES; MAGNETIC RESONANCE (MR) CINE STUDY; MAGNETIC RESONANCE (MR); TECHNOLOGY;
D O I
10.1148/radiology.178.2.1987592
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Six healthy volunteers and three patients with cardiac anomalies were studied in a comparison of segmented turboFLASH (fast low-angle shot) cine, a method of magnetic resonance imaging that permits an entire series of high-resolution cine images to be obtained in one breath hold, with standard cine. Segmented turboFLASH uses a gradient-echo sequence designed for short imaging times in combination with a segmented data acquisition method. Presaturation pulses were applied to eliminate the blood pool signal; the signal-to-noise ratio was assessed with a phantom. Standard hardware and image reconstruction methods were used. The breath-hold images consistently showed reduced ghosting and blurring from respiration. Because a very short echo time was used, segmented turboFLASH was relatively insensitive to dephasing caused by local field disturbances or flow. The authors conclude that, by reducing imaging times and eliminating respiratory artifact, segmented turboFLASH can be useful for performing cine studies of the heart and great vessels.
引用
收藏
页码:357 / 360
页数:4
相关论文
共 10 条
  • [1] 1ST-PASS CARDIAC PERFUSION - EVALUATION WITH ULTRAFAST MR IMAGING
    ATKINSON, DJ
    BURSTEIN, D
    EDELMAN, RR
    [J]. RADIOLOGY, 1990, 174 (03) : 757 - 762
  • [2] ATKINSON DJ, 1988, 1988 SOC MAGN RES ME, P137
  • [3] ECHOCARDIOGRAPHIC-LIKE ANGLED VIEWS OF THE HEART BY MR IMAGING
    BURBANK, F
    PARISH, D
    WEXLER, L
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1988, 12 (02) : 181 - 195
  • [4] SEGMENTED TURBOFLASH - METHOD FOR BREATH-HOLD MR IMAGING OF THE LIVER WITH FLEXIBLE CONTRAST
    EDELMAN, RR
    WALLNER, B
    SINGER, A
    ATKINSON, DJ
    SAINI, S
    [J]. RADIOLOGY, 1990, 177 (02) : 515 - 521
  • [5] 0.3-SECOND FLASH MRI OF THE HUMAN-HEART
    FRAHM, J
    MERBOLDT, KD
    BRUHN, H
    GYNGELL, ML
    HANICKE, W
    CHIEN, D
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1990, 13 (01) : 150 - 157
  • [6] HAASE A, 1989, 1989 SOC MAGN RES ME, P56
  • [7] FUNCTIONAL-EVALUATION OF THE HEART WITH MAGNETIC-RESONANCE IMAGING
    HIGGINS, CB
    HOLT, W
    PFLUGFELDER, P
    SECHTEM, U
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1988, 6 (02) : 121 - 139
  • [8] THE VALUE OF CINE NUCLEAR MAGNETIC-RESONANCE IMAGING FOR ASSESSING REGIONAL VENTRICULAR-FUNCTION
    LOTAN, CS
    CRANNEY, GB
    BOUCHARD, A
    BITTNER, V
    POHOST, GM
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (07) : 1721 - 1729
  • [9] RAPID BIOMEDICAL IMAGING BY NMR
    ORDIDGE, RJ
    MANSFIELD, P
    COUPLAND, RE
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1981, 54 (646) : 850 - 855
  • [10] WHITE RD, 1989, 1989 SOC MAGN RES ME, P58