OPTIMIZATION OF AN HPLC PEROXYOXALATE CHEMI-LUMINESCENCE DETECTION SYSTEM FOR SOME DANSYL AMINO-ACIDS

被引:14
作者
BAEYENS, W
BRUGGEMAN, J
DEWAELE, C
LIN, B
IMAI, K
机构
[1] STATE UNIV GHENT, FAC SCI, ORGAN CHEM LAB, B-9000 GHENT, BELGIUM
[2] UNIV TOKYO, BRANCH HOSP PHARM, BUNKYO KU, TOKYO 112, JAPAN
来源
JOURNAL OF BIOLUMINESCENCE AND CHEMILUMINESCENCE | 1990年 / 5卷 / 01期
关键词
D O I
10.1002/bio.1170050105
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
Bis(2,4,6-trichlorophenyl) oxalate (TCPO)-hydrogen-peroxide-generated chemiluminescence (CL) of four dansyl amino acids has been used as a model system for the optimization of a detection system in reversed-phase high-performance liquid chromatography. Dansylated alanine, glutamic acid, methionine, and norleucine were subjected to peroxyoxalate induced CL in a static system and in a flow system under various conditions with respect to TCPO (ethyl acetate) and hydrogen peroxide (acetone) concentrations, solvent composition and flow, using a two-pump or a one-pump post-column reagent system. From the CL-decay curve, the influence on the emission signal from the total flow rate in the detector was investigated. Special attention was focused on the mixing of the LC eluate and the reagent in order to combine an efficient collection of the emitted light using a 74 microliter flow cell (originally 10 microliters in the fluorescence detector) with minimal extra column band broadening. Therefore, a capillary fused-silica tubing of about 100 microns i.d. was inserted against the end-frit of the column and brought through a mixing tee, in which the solutions of TCPO and hydrogen peroxide were added. The column end tubing ended in the flow cell and the LC eluate and the reagents were mixed when entering the flow-cell. Average detection limits (S/N = 2) of 200 fmol injected dansylated amino acid could be reached. A comparison is made between the use of TCPO and DNPO (bis (2, 4-dinitrophenyl) oxalate).
引用
收藏
页码:13 / 23
页数:11
相关论文
共 34 条
[1]   DETERMINATION OF MORPHINE IN BODY-FLUIDS BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH CHEMILUMINESCENCE DETECTION [J].
ABBOTT, RW ;
TOWNSHEND, A ;
GILL, R .
ANALYST, 1987, 112 (04) :397-406
[2]  
BAEYENS W, 1987, CHEM MAGAZINE, V6, P15
[3]  
BAEYENS W, 1987, CHEM MAGAZINE, V13, P17
[4]   FLOW-INJECTION ANALYSIS FOR MONITORING CHEMI-LUMINESCENT REACTIONS [J].
BURGUERA, JL ;
TOWNSHEND, A ;
GREENFIELD, S .
ANALYTICA CHIMICA ACTA, 1980, 114 (FEB) :209-214
[5]   DETERMINATION OF AMINES BY USING A CHEMILUMINESCENT REACTION IN SOLUTION [J].
BURGUERA, JL ;
TOWNSHEND, A .
TALANTA, 1979, 26 (09) :795-798
[6]   OPTIMIZATION OF A PEROXYOXALATE CHEMI-LUMINESCENCE DETECTION SYSTEM FOR THE LIQUID-CHROMATOGRAPHIC DETERMINATION OF FLUORESCENT COMPOUNDS [J].
DEJONG, GJ ;
LAMMERS, N ;
SPRUIT, FJ ;
BRINKMAN, UAT ;
FREI, RW .
CHROMATOGRAPHIA, 1984, 18 (03) :129-133
[7]   LOW-DISPERSION CHEMILUMINESCENCE DETECTION FOR PACKED CAPILLARY LIQUID-CHROMATOGRAPHY [J].
DEJONG, GJ ;
LAMMERS, N ;
SPRUIT, FJ ;
DEWAELE, C ;
VERZELE, M .
ANALYTICAL CHEMISTRY, 1987, 59 (10) :1458-1461
[8]   ANALYTICAL IMPLICATIONS OF THE HALF-LIFE OF THE CHEMILUMINESCENCE SIGNAL IN THE PEROXYOXALATE DETECTION SYSTEM FOR LIQUID-CHROMATOGRAPHY [J].
DEJONG, GJ ;
LAMMERS, N ;
SPRUIT, FJ ;
FREI, RW ;
BRINKMAN, UAT .
JOURNAL OF CHROMATOGRAPHY, 1986, 353 :249-257
[9]   INVESTIGATION OF INSTRUMENT PARAMETERS FOR THE APPLICATION OF PEROXYOXALATE CHEMI-LUMINESCENCE DETECTION TO MICROBORE CHROMATOGRAPHY [J].
GRAYESKI, ML ;
WEBER, AJ .
ANALYTICAL LETTERS PART A-CHEMICAL ANALYSIS, 1984, 17 (13) :1539-1552
[10]  
GRAYESKI ML, 1987, ANAL CHEM, V59, pA1243