DETERMINATION OF IMMUNOGLOBULIN-M CONCENTRATION BY PIEZOELECTRIC CRYSTAL IMMUNOBIOSENSOR COATED WITH PROTAMINE

被引:21
作者
SURI, CR [1 ]
RAJE, M [1 ]
MISHRA, GC [1 ]
机构
[1] INST MICROBIAL TECHNOL,SECTOR 39-A,CHANDIGARH 160014,INDIA
关键词
IMMUNOBIOSENSOR; AT-CUT PIEZOELECTRIC CRYSTAL; OSCILLATOR; HYDROXYLIC GROUPS; TRESYL CHLORIDE; IMMUNOGLOBULINS;
D O I
10.1016/0956-5663(94)80031-6
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
In the present study, the specific binding between protamine and immunoglobulin M (IgM) has been exploited to construct a piezoelectric crystal based immunobiosensor for the determination of concentration of IgM. The system consisted of highly stable IC based oscillator, 8-digit frequency counter and modified piezoelectric crystal device. The crystal surface was physically modified and chemically treated (refluxed) with strong acid to produce stable hydroxylic groups of silicon oxide. This modified surface reacted strongly with coupling reagents for binding of protein molecules. The protamine was immobilized by using either gamma-aminopropyltriethoxy silane (gamma-APTES) or 2.2.2-trifluoroethanesulfonyl chloride (tresyl chloride). Scanning electron microscope images of piezo crystal revealed that tresyl activated surface presented more surface area for binding than gamma-APTES modified surface and showed better sensitivity. This immobilization technique also improved the reproducibility and long term stability of the detection system. Using the system described, the IgM concentration up to the level of 10 ng/ml could be detected without interference of IgG.
引用
收藏
页码:325 / 332
页数:8
相关论文
共 28 条
[1]   A CONCANAVALIN A-COATED PIEZOELECTRIC CRYSTAL BIOSENSOR [J].
BARNES, C ;
DSILVA, C ;
JONES, JP ;
LEWIS, TJ .
SENSORS AND ACTUATORS B-CHEMICAL, 1991, 3 (04) :295-304
[2]   DETECTION OF TRACE CONCENTRATIONS OF GASES WITH COATED PIEZOELECTRIC QUARTZ CRYSTALS [J].
BEITNES, H ;
SCHRODER, K .
ANALYTICA CHIMICA ACTA, 1984, 158 (01) :57-65
[3]  
DAWSON RM, 1986, DATA BIOCH RES, P466
[4]   DETECTION OF GALACTOSYLTRANSFERASE USING CHEMICALLY MODIFIED PIEZOELECTRIC QUARTZ [J].
GRANDE, LH ;
GEREN, CR ;
PAUL, DW .
SENSORS AND ACTUATORS, 1988, 14 (04) :387-403
[5]   A PIEZOELECTRIC IMMUNOBIOSENSOR FOR ATRAZINE IN DRINKING-WATER [J].
GUILBAULT, GG ;
HOCK, B ;
SCHMID, R .
BIOSENSORS & BIOELECTRONICS, 1992, 7 (06) :411-419
[6]   IMMOBILIZATION METHODS FOR PIEZOELECTRIC BIOSENSORS [J].
GUILBAULT, GG ;
LUONG, JHT ;
PRUSAKSOCHACZEWSKI, E .
BIO-TECHNOLOGY, 1989, 7 (04) :349-351
[7]  
HUDSON L, 1980, PRACTICAL IMMUNOLOGY, P220
[8]  
IZUI S, 1984, J IMMUNOL, V133, P227
[9]  
JIMBO Y, 1990, BIOINSTRUMENTATION R, P466
[10]   INDIRECT DOUBLE SANDWICH ELISA FOR THE SPECIFIC AND QUANTITATIVE MEASUREMENT OF MOUSE IGM, IGA AND IGG SUBCLASSES [J].
KLEINSCHNEEGANS, AS ;
GAVERIAUX, C ;
FONTENEAU, P ;
LOOR, F .
JOURNAL OF IMMUNOLOGICAL METHODS, 1989, 119 (01) :117-125