CRYOFOCUSING INLET WITH REVERSE FLOW SAMPLE COLLECTION FOR GAS-CHROMATOGRAPHY

被引:59
作者
KLEMP, MA [1 ]
AKARD, ML [1 ]
SACKS, RD [1 ]
机构
[1] UNIV MICHIGAN,DEPT CHEM,ANN ARBOR,MI 48109
关键词
D O I
10.1021/ac00066a020
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An inlet system for high-speed gas chromatography is characterized. Sample is cryofocused in a bare metal capillary tube and released by a fast temperature increase caused by an electric current. The gas flow direction through the capillary tube is reversed after sample collection so that the sample injection onto the capillary column for separation occurs from the downstream end of the metal tube. The system offers narrower initial band widths, elimination of memory effects, and reduced thermal decomposition during reinjection relative to previously described cryofocusing inlet systems. Sample vapor can be collected and cryofocused from any source pressure greater than approximately 20 Torr. A configuration for direct air sampling is described.
引用
收藏
页码:2516 / 2521
页数:6
相关论文
共 23 条
[1]   THE USE OF COANDA WALL ATTACHMENT FLUIDIC SWITCHES AS GAS-CHROMATOGRAPHIC VALVES [J].
ANNINO, R ;
LEONE, J .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1982, 20 (01) :19-26
[2]   ELECTRICALLY HEATED COLD TRAP INLET SYSTEM FOR HIGH-SPEED GAS-CHROMATOGRAPHY [J].
EWELS, BA ;
SACKS, RD .
ANALYTICAL CHEMISTRY, 1985, 57 (14) :2774-2779
[3]   STUDY IN HIGH-SPEED GAS-CHROMATOGRAPHY .1. INJECTIONS OF NARROW SAMPLE PLUGS [J].
GASPAR, G ;
ARPINO, P ;
GUIOCHON, G .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1977, 15 (07) :256-261
[4]   PLATE HEIGHT IN GAS CHROMATOGRAPHY [J].
GIDDINGS, JC ;
SEAGER, SL ;
STUCKI, LR ;
STEWART, GH .
ANALYTICAL CHEMISTRY, 1960, 32 (07) :867-870
[5]   HOW EFFICIENT ARE CAPILLARY COLD TRAPS [J].
GRAYDON, JW ;
GROB, K .
JOURNAL OF CHROMATOGRAPHY, 1983, 254 (JAN) :265-269
[6]   RAPID EVAPORATION OF CONDENSED GAS-CHROMATOGRAPHIC FRACTIONS [J].
HOPKINS, BJ ;
PRETORIUS, V .
JOURNAL OF CHROMATOGRAPHY, 1978, 158 (OCT) :465-469
[7]   SYSTEMATIC-ERRORS OCCURRING WITH THE USE OF GAS-SAMPLING LOOP INJECTORS IN GAS-CHROMATOGRAPHY [J].
JONSSON, JA ;
VEJROSTA, J ;
NOVAK, J .
JOURNAL OF CHROMATOGRAPHY, 1982, 236 (02) :307-312
[8]   VACUUM BACKFLUSH RECYCLE SYSTEM FOR GC ANALYSIS OF HIGH-PURITY SOLVENTS [J].
KLEMP, M ;
SACKS, RD .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1991, 29 (03) :114-121
[9]   STRATEGIES FOR RAPID GC IMPURITY ANALYSIS OF HIGH-PURITY SOLVENTS .1. INSTRUMENTATION AND PROCEDURES [J].
KLEMP, M ;
SACKS, R .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1991, 29 (06) :243-247
[10]  
KLEMP M, 1991, HRC-J HIGH RES CHROM, V14, P235