PHOSPHOR PLATE COMPUTED RADIOGRAPHY - RESPONSE TO VARIATION IN MAS AT FIXED KVP IN AN ANIMAL-MODEL - POTENTIAL ROLE IN NEONATAL IMAGING

被引:9
作者
BRODERICK, NJ
LONG, B
DREESEN, RG
COHEN, MD
CORY, DA
KATZ, BP
KALASINSKI, LA
机构
[1] INDIANA UNIV,SCH MED,REGENSTRIEF HLTH CTR,DEPT MED,INDIANAPOLIS,IN 46202
[2] INDIANA UNIV,SCH MED,RILEY HOSP CHILDREN,DEPT RADIOL,INDIANAPOLIS,IN 46202
关键词
D O I
10.1016/S0009-9260(05)81212-7
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Computed digital radiography offers a number of potential advantages over conventional film-screen imaging. To determine the applicability of these to portable neonatal examinations we performed a controlled prospective study comparing conventional 200 speed film-screen with a computed radiography unit in an animal model. Images acceptable for diagnostic purposes were obtained on the digital system over an exposure range many orders of magnitude greater than was possible on the film-screen system. Digital imaging offers potential for elimination of repeat examinations due to suboptimal exposure factors, and for reduction in radiation dose to the patient. We believe that computed digital radiography should be particularly suited to portable neonatal imaging.
引用
收藏
页码:39 / 45
页数:7
相关论文
共 20 条
[1]   DIGITAL AND CONVENTIONAL CHEST IMAGING - A MODIFIED ROC STUDY OF OBSERVER PERFORMANCE USING SIMULATED NODULES [J].
CHAKRABORTY, DP ;
BREATNACH, ES ;
YESTER, MV ;
SOTO, B ;
BARNES, GT ;
FRASER, RG .
RADIOLOGY, 1986, 158 (01) :35-39
[2]  
DOI K, 1986, RADIOLOGY, V161, P513, DOI 10.1148/radiology.161.2.3763924
[3]  
DREW PG, 1988, DIAGNOSTIC IMAGINGX, V5, P47
[4]   EXCRETORY UROGRAPHY USING COMPUTED RADIOGRAPHY [J].
FAJARDO, LL ;
HILLMAN, BJ ;
HUNTER, TB ;
CLAYPOOL, HR ;
WESTERMAN, BR ;
MOCKBEE, B .
RADIOLOGY, 1987, 162 (02) :345-351
[5]  
FRASER RG, 1983, RADIOLOGY, V148, P1
[6]   DIGITAL IMAGING OF THE CHEST [J].
FRASER, RG ;
SANDERS, C ;
BARNES, GT ;
MACMAHON, H ;
GIGER, ML ;
DOI, K ;
TEMPLETON, AW ;
COX, GG ;
DWYER, SJ ;
MERRITT, CRB ;
JONES, JP .
RADIOLOGY, 1989, 171 (02) :297-307
[7]   STORAGE PHOSPHOR RADIOGRAPHS VS CONVENTIONAL FILMS - INTERPRETERS PERCEPTIONS OF DIAGNOSTIC QUALITY [J].
FUHRMAN, CR ;
GUR, D ;
GOOD, B ;
ROCKETTE, H ;
COOPERSTEIN, LA ;
FEIST, JH .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1988, 150 (05) :1011-1014
[8]   DIGITAL RADIOGRAPHY OF THE CHEST - PROMISES AND PROBLEMS [J].
GOODMAN, LR ;
WILSON, CR ;
FOLEY, WD .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1988, 150 (06) :1241-1252
[9]  
KIRKS DR, 1984, PRACTICAL PEDIATRIC, P7
[10]   LOW-DOSE DIGITAL COMPUTED RADIOGRAPHY IN PEDIATRIC CHEST IMAGING [J].
KOGUTT, MS ;
JONES, JP ;
PERKINS, DD .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1988, 151 (04) :775-779