MULTICENTER STUDY OF A PORTABLE, HAND-SIZE, COLORIMETRIC END-TIDAL CARBON-DIOXIDE DETECTION DEVICE

被引:93
作者
ORNATO, JP
SHIPLEY, JB
RACHT, EM
SLOVIS, CM
WRENN, KD
PEPE, PE
ALMEIDA, SL
GINGER, VF
FOTRE, TV
机构
[1] Medical College of Virginia, Virginia Commonwealth University, Richmond
[2] Strong Memorial Hospital, University of Rochester, Rochester, NY
[3] Baylor College of Medicine, City of Houston Center for Resuscitation, Emergency Medical Services, San Mateo, CA
[4] City of Houston Emergency Medical Services, San Mateo, CA
[5] San Mateo Emergency Medical Services Agency, San Mateo, CA
关键词
END-TIDAL CO2 DETECTION;
D O I
10.1016/S0196-0644(05)82517-X
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Study objectives: To evaluate continuous, semiquantitative end-tidal carbon dioxide (ETCO2) monitoring in the prehospital and emergency department setting for confirming proper endotracheal tube placement and assessing prognosis and blood flow during CPR. Type of participants: Adult patients were included if an endotracheal tube was inserted by prehospital care providers or emergency physicians for cardiac arrest, respiratory arrest, respiratory insufficiency, or airway protection. Design and interventions: A small, portable, colorimetric ETCO2 detector was attached to the endotracheal tube immediately after each attempted endotracheal tube insertion. The color of the detector membrane was noted at the seventh breath following intubation. The color also was noted and recorded if there was return of spontaneous circulation (defined as a palpable pulse) immediately prior to and following conversion from manual to mechanical CPR. Survival to hospital admission was used as an end point to assess the prognostic value of the initial ETCO2 reading. Main results: A total of 227 patients (144 with cardiopulmonary arrest) were studied. In the 83 patients intubated but not in cardiopulmonary arrest, a reading on the ETCO2 detector signifying more than 0.5% ETCO2 was 100% sensitive and 93% specific in detecting proper endotracheal tube placement (100% specific with the endotracheal tube cuff inflated). In cardiac arrest patients, a longer period of estimated arrest appeared to be associated with a lower ETCO2 detector reading. A reading signifying more than 0.5% ETCO2 was 69% sensitive and 100% specific in detecting proper endotracheal tube placement. After proper endotracheal tube placement, all cardiac arrest patients who survived to hospital admission had an initial ETCO2 measurement signifying more than 0.5% ETCO2. Return of spontaneous circulation was usually accompanied by an improved ETCO2 value. Mechanical CPR always produced an ETCO2 value that was as high or higher than that produced by manual CPR. Conclusion: The colorimetric ETCO2 device is highly accurate for confirming endotracheal tube position in nonarrest patients. In cardiac arrest patients, a reading signifying more than 0.5% ETCO2 confirms correct endotracheal tube placement, while a value signifying less than 0.5% ETCO2 during resuscitation suggests that something is wrong (eg, esophageal intubation, inadequate circulatory flow, prolonged downtime interval, hypothermia, or significant ventilation/perfusion mismatch).
引用
收藏
页码:518 / 523
页数:6
相关论文
共 26 条
[1]  
Smallhout, Kalenda, An Atlas of Capnography, 1, pp. 28-33, (1975)
[2]  
Garnett, Ornato, Gonzalez, Et al., End-tidal carbon dioxide measurement during cardiopulmonary resuscitation, JAMA, 257, pp. 512-515, (1987)
[3]  
Falk, Rackow, Weil, End-tidal carbon dioxide concentration during cardiopulmonary resuscitation, N Engl J Med, 318, pp. 607-611, (1988)
[4]  
Gudipati, Weil, Bisera, Et al., Expired carbon dioxide: A noninvasive monitor of cardiopulmonary resuscitation, Circulation, 77, pp. 234-239, (1988)
[5]  
Weil, Bisera, Trevino, Et al., Cardiac output and end-tidal carbon dioxide, Crit Care Med, 13, pp. 907-909, (1985)
[6]  
Kalenda, The capnogram as a guide to the efficacy of cardiac massage, Resuscitation, 6, pp. 259-263, (1978)
[7]  
Sanders, Atlas, Ewy, Et al., Expired pCO<sub>2</sub> as an index of coronary perfusion pressure, Am J Emerg Med, 3, pp. 147-149, (1985)
[8]  
Kern, Sanders, Voorhees, Et al., Changes in expired end-tidal carbon dioxide during cardiopulmonary resuscitation in dogs: A prognostic guide for resuscitation efforts, JAm Coll Cardiol, 13, pp. 1184-1189, (1989)
[9]  
Sanders, Ewy, Bragg, Et al., Expired PCO<sub>2</sub> as a prognostic indicator of successful resuscitation from cardiac arrest, Ann Emerg Med, 14, pp. 948-952, (1985)
[10]  
Sanders, Kern, Otto, Et al., End-tidal carbon dioxide monitoring during cardiopulmonary resuscitation: A prognostic indicator for survival, JAMA, 262, pp. 1347-1351, (1989)