A LOW-NOISE DEMULTIPLEXING SYSTEM FOR ACTIVE MULTICHANNEL MICROELECTRODE ARRAYS

被引:18
作者
JI, J
NAJAFI, K
WISE, KD
机构
[1] Center for Integrated Sensors and Circuits, Department of Electrical Engineering and Computer Science, University of Michigan, Ann Arbor
关键词
PLANAR;
D O I
10.1109/10.68212
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper reports a low-noise demultiplexing system capable of reconstructing multichannel single-unit neural signals derived from multiplexed microelectrode arrays. The overall multiplexing-demultiplexing system realizes ten channels, a per-channel gain of 68 dB, a bandwidth from 100 Hz to 6 kHz, and an equivalent noise level (referred to the probe input) of 13-mu-V rms. It provides for signaling over the power supply to allow control of on-chip probe functions such as self-testing. The interchannel crosstalk is less than 3%, and switching noise is suppressed by blanking the transition intervals. The 200 kHz probe sample clock is tracked automatically over a range from 150 to 250 kHz. Neural signals as low as 20-mu-V (typically 640-mu-V at the demultiplexing system input) can be reconstructed. The overall system organization is compatible with the demultiplexing of as many as 40 time-multiplexed electrode channels from a single probe data line.
引用
收藏
页码:75 / 81
页数:7
相关论文
共 8 条
[1]  
BNAJAFI K, 1985, EEE T ELECTRON DEVIC, V22, P1206
[2]   PERFORMANCE OF PLANAR MULTISITE MICROPROBES IN RECORDING EXTRACELLULAR SINGLE-UNIT INTRACORTICAL ACTIVITY [J].
DRAKE, KL ;
WISE, KD ;
FARRAYE, J ;
ANDERSON, DJ ;
BEMENT, SL .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (09) :719-732
[3]  
HETKE JF, 1989, 5 INT C SOL STAT SEN, P277
[4]  
JI J, 1989, 5TH INT C SOL STAT S, P195
[5]   A PRACTICAL 2J CHANNEL MICROELECTRODE FOR NEURAL RECORDING INVIVO [J].
KUPERSTEIN, M ;
WHITTINGTON, DA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1981, 28 (03) :288-293
[6]   AN IMPLANTABLE MULTIELECTRODE ARRAY WITH ON-CHIP SIGNAL-PROCESSING [J].
NAJAFI, K ;
WISE, KD .
IEEE JOURNAL OF SOLID-STATE CIRCUITS, 1986, 21 (06) :1035-1044
[7]   INTEGRATION OF MULTI-MICROELECTRODE AND INTERFACE CIRCUITS BY SILICON PLANAR AND 3-DIMENSIONAL FABRICATION TECHNOLOGY [J].
TAKAHASHI, K ;
MATSUO, T .
SENSORS AND ACTUATORS, 1984, 5 (01) :89-99
[8]   AN INTEGRATED-CIRCUIT APPROACH TO EXTRACELLULAR MICROELECTRODES [J].
WISE, KD ;
ANGELL, JB ;
STARR, A .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1970, BM17 (03) :238-+