IMPROVED CALIBRATION OF A LARGE OPEN-ENDED COAXIAL PROBE FOR DIELECTRIC MEASUREMENTS

被引:30
作者
OTTO, GP
CHEW, WC
机构
关键词
D O I
10.1109/19.85345
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The accurate calibration of open-ended coaxial waveguides can be difficult because standard matched loads over a broad frequency range cannot be easily attached without connectors that will affect the measurements. An alternative termination is the open-ended probe immersed in a material with known permittivity. But this termination is often inaccurate due to the multiple sources of uncertainty in the reference permittivity. An improved calibration is obtained by measuring an open, a short, and a short-cavity. This calibration is more accurate than measurements of reference permittivities because it depends only upon the geometry of the probe. The calibration was tested by comparing theoretical and experimental values of the complex permittivity for common liquids, and the discrepancy was seen to be small.
引用
收藏
页码:742 / 746
页数:5
相关论文
共 22 条
[1]  
Abramowitz M., 1972, HDB MATH FUNCTIONS, P484
[2]   ANALYSIS OF AN OPEN-ENDED COAXIAL LINE SENSOR IN LAYERED DIELECTRICS [J].
ANDERSON, LS ;
GAJDA, GB ;
STUCHLY, SS .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1986, 35 (01) :13-18
[3]   MEASUREMENT OF RADIO-FREQUENCY PERMITTIVITY OF BIOLOGICAL TISSUES WITH AN OPEN-ENDED COAXIAL LINE .1. [J].
ATHEY, TW ;
STUCHLY, MA ;
STUCHLY, SS .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1982, 30 (01) :82-86
[4]   EFFECT OF FINITE SIZE OF GROUND PLANE ON THE IMPEDANCE OF A MONOPOLE IMMERSED IN A LOSSY MEDIUM [J].
BAHL, IJ ;
STUCHLY, SS .
ELECTRONICS LETTERS, 1979, 15 (22) :728-729
[5]  
BRADY MM, 1981, IEEE T BIO-MED ENG, V28, P305, DOI 10.1109/TBME.1981.324707
[6]  
BUCKLEY F, 1958, NBS598 CIRC
[7]   INVIVO PROBE MEASUREMENT TECHNIQUE FOR DETERMINING DIELECTRIC-PROPERTIES AT VHF THROUGH MICROWAVE-FREQUENCIES [J].
BURDETTE, EC ;
CAIN, FL ;
SEALS, J .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1980, 28 (04) :414-427
[8]   DESIGN AND CALIBRATION OF A LARGE BROAD-BAND DIELECTRIC MEASUREMENT CELL [J].
CHEW, WC ;
OLP, KJ ;
OTTO, GP .
IEEE TRANSACTIONS ON GEOSCIENCE AND REMOTE SENSING, 1991, 29 (01) :42-47
[9]   NUMERICAL-ANALYSIS OF OPEN-ENDED CO-AXIAL LINES [J].
GAJDA, GB ;
STUCHLY, SS .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1983, 31 (05) :380-384
[10]  
HILDEBRAND FB, 1976, ADV CALCULUS APPL, P226