CRITICAL ANALYSIS OF THE ACCURACY OF A COMBINED TEST (SE-75-HCAT AND CO-58 B-12) OF ILEAL FUNCTION USING A WHOLE-BODY COUNTER

被引:6
作者
SMITH, T [1 ]
BJARNASON, I [1 ]
HESP, R [1 ]
机构
[1] MRC,DIV CLIN SCI,HARROW HA1 3UJ,MIDDX,ENGLAND
来源
CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT | 1987年 / 8卷 / 02期
关键词
BIOMEDICAL EQUIPMENT - Radionuclides - NUCLEAR MEDICINE - RADIOACTIVE MATERIALS;
D O I
10.1088/0143-0815/8/2/006
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The accuracy of a combined test (**7**5Se-HCAT plus **5**8Co-B//1//2) of ileal function by whole-body counting has been assessed with particular attention being paid to the reliability of the 100% retention measurements, the accuracy of the crossover of **5**8Co into the **7**5Se channel and the effects of variable counting geometry. A linear relationship between **5**8Co crossover and body weight improved the accuracy of determining the crossover value for individual patients. The main sources of error were due to problems of variable counting geometry but there was very good agreement between **7**5Se and **5**8Co retentions measured by whole body counting and by excreta analysis which suggested that the overall error of the test was unlikely to exceed plus or minus 10%. A protocol is presented for use of the combined test on an outpatient basis and includes retention measurements at 4, 7, and 10 d post administration.
引用
收藏
页码:159 / 171
页数:13
相关论文
共 14 条
  • [1] BJARNASON I, 1987, IN PRESS GASTROENTER
  • [2] NEW METHOD FOR THE DETERMINATION OF BILE-ACID TURNOVER USING SE-75-HOMOCHOLIC ACID TAURINE
    DELHEZ, H
    VANDENBERG, JWQ
    VANBLANKENSTEIN, M
    MEERWALDT, JH
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1982, 7 (06): : 269 - 271
  • [3] SEHCAT ABSORPTION - A SIMPLE TEST OF ILEAL DYSFUNCTION
    FAGAN, EA
    CHADWICK, VS
    BAIRD, IM
    [J]. DIGESTION, 1983, 26 (03) : 159 - 165
  • [4] USE OF A GAMMA-LABELED BILE-ACID (SE-75HCAT) AS A TEST OF ILEAL FUNCTION - METHODS OF IMPROVING ACCURACY
    FERRARIS, R
    JAZRAWI, R
    BRIDGES, C
    NORTHFIELD, TC
    [J]. GASTROENTEROLOGY, 1986, 90 (05) : 1129 - 1136
  • [5] A COMPARISON BETWEEN THE USE OF A SHADOW SHIELD WHOLE-BODY COUNTER AND AN UNCOLLIMATED GAMMA-CAMERA IN THE ASSESSMENT OF THE 7-DAY RETENTION OF SEHCAT
    HAMES, TK
    FLEMING, JS
    PHILLIPS, G
    HOLDSTOCK, G
    SMITH, CL
    HOWLETT, PJ
    ACKERY, D
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1984, 57 (679) : 581 - 584
  • [6] APPLICATION OF DIFFERENT COUNTING SYSTEMS IN MEASURING LOCALIZED RETENTION IN THE ENTERO-HEPATIC CIRCULATION
    HINTON, PJ
    TAYLOR, DN
    WYKE, RJ
    MCINTOSH, JA
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1984, 57 (681) : 799 - 801
  • [7] HOLDSTOCK G, 1985, EUR J NUCL MED, V10, P528
  • [8] MERRICK MV, 1982, J NUCL MED, V23, P126
  • [9] NYHLIN H, 1983, GASTROENTEROLOGY, V84, P63
  • [10] SCHROTH HJ, 1985, EUR J NUCL MED, V10, P455