A NEW TRANSDUCER-TIPPED FIBER OPTIC CATHETER FOR MEASURING INTRAMUSCULAR PRESSURES

被引:23
作者
CRENSHAW, AG
STYF, JR
MUBARAK, SJ
HARGENS, AR
机构
[1] VET ADM MED CTR,DIV ORTHOPAED & REHABIL,SAN DIEGO,CA 92161
[2] UNIV CALIF SAN DIEGO,LA JOLLA,CA 92093
[3] NASA,AMES RES CTR,MOFFETT FIELD,CA 94035
关键词
Fiber optics; Hydrostatic artifact; Intramuscular pressure; Transducer‐tipped catheter;
D O I
10.1002/jor.1100080318
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Laboratory and clinical tests were used to determine the efficacy of a new fiber optic “transducer‐tipped” catheter for measuring intramuscular pressures. When pressures ranging from 0 to 250 mm Hg were applied by a mercury manometer, the fiber optic system accurtely recorded the pressures. In addition, the fiber optic system showed long‐term stability by accurately recording the manometer pressure over a 3 day period. The Slit system showed a higher magnitude of hydrostatic pressure artifacts with catheter tip movement as compared to the fiber optic system. The two catheters showed no difference when measuring pressures in pig muscle at rest or when being compressed throughout a range of 0 to 250 mm Hg. In human volunteers, both catheters measured essentially equal pressures at rest, during venous stasis, and during a combination of venous stasis and compression. For long‐term assessment, the Slit system required as many as three saline flushes, whereas the fiber optic system measured pressures continuously without manipulation. We conclude that the fiber optic system is as accurate as the Slit catheter for measuring tissue fluid pressures at rest. In addition, the fiber optic system offers distinct advantages over conventional fluid‐filled systems for measuring intramuscular pressures due to a lack of hydrostatic pressure artifacts caused by limb position and to the lack of flushing for long‐term measurements. Copyright © 1990 Orthopaedic Research Society
引用
收藏
页码:464 / 468
页数:5
相关论文
共 12 条
[1]   RE-EVALUATION OF NEEDLE METHOD FOR MEASURING INTERSTITIAL FLUID PRESSURE [J].
BRACE, RA ;
GUYTON, AC ;
TAYLOR, AE .
AMERICAN JOURNAL OF PHYSIOLOGY, 1975, 229 (03) :603-607
[2]  
CORNO AF, 1987, J THORAC CARDIOV SUR, V93, P163
[3]   PRESSURE-VOLUME CHARACTERISTICS OF INTERSTITIAL FLUID SPACE IN SKELETAL-MUSCLE OF CAT [J].
ELIASSEN, E ;
FOLKOW, B ;
HILTON, SM ;
OBERG, B ;
RIPPE, B .
ACTA PHYSIOLOGICA SCANDINAVICA, 1974, 90 (03) :583-593
[4]  
Guyton AC, 1981, TISSUE FLUID PRESSUR, P11
[5]  
MATSEN FA, 1976, SURGERY, V79, P702
[6]  
MCDERMOTT AGP, 1981, AM J SPORTS MED, V10, P83
[7]  
MUBARAK SJ, 1981, COMPARTMENT SYNDROME
[8]   CONTINUOUS MONITORING OF INTRACRANIAL-PRESSURE WITH A MINIATURIZED FIBEROPTIC DEVICE [J].
OSTRUP, RC ;
LUERSSEN, TG ;
MARSHALL, LF ;
ZORNOW, MH .
JOURNAL OF NEUROSURGERY, 1987, 67 (02) :206-209
[9]   INTRAARTICULAR PRESSURE DURING CONTINUOUS PASSIVE MOTION OF THE HUMAN KNEE [J].
PEDOWITZ, RA ;
GERSHUNI, DH ;
CRENSHAW, AG ;
PETRAS, SL ;
DANZIG, LA ;
HARGENS, AR .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1989, 7 (04) :530-537
[10]  
RORABECK CH, 1981, J TRAUMA, V21, P446