ATRIAL ARRHYTHMIA MANAGEMENT WITH SENSOR CONTROLLED ATRIAL REFRACTORY PERIOD AND AUTOMATIC-MODE SWITCHING IN PATIENTS WITH MINUTE VENTILATION SENSING DUAL CHAMBER RATE ADAPTIVE PACEMAKERS

被引:37
作者
LAU, CP
TAI, YT
FONG, PC
LI, JPS
CHUNG, FLW
机构
[1] Division of Cardiology, Department of Medicine, University of Hong Kong, Queen Mary Hospital
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1992年 / 15卷 / 10期
关键词
ATRIAL ARRHYTHMIAS; PACEMAKER MEDIATED TACHYCARDIAS; RATE ADAPTIVE DUAL CHAMBER PACING; SENSORS;
D O I
10.1111/j.1540-8159.1992.tb02924.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Although a long postventricular atrial refractory period (PVARP) may prevent the occurrence of pacemaker mediated tachycardias and inadvertent tracking of atrial arrhythmias in dual chamber (DDD) pacing, the maximum upper rate will necessarily be compromised. We tested the feasibility of using minute ventilation sensing in a dual chamber rate adaptive pacemaker (DDDR) to shorten the PVARP during exercise in 13 patients with bradycardias (resting PVARP = 463 +/- 29 msec) to avoid premature upper rate behavior. Graded treadmill exercise tests in the DDD and DDDR modes at this PVARP resulted in maximum ventricular rates of 98 +/- 8 and 142 +/- 3 beats/min, respectively (P < 0.0001), due to chronotropic incompetence and upper rate limitation in the DDD mode, both circumvened with the use of sensor. In order to simulate atrial arrhythmias, chest wall stimulation was applied for 30 seconds at a rate of 250 beats/min at a mean unipolar atrial sensitivity of 0.82 mV. Irregular ventricular responses occurred in the DDD mode (the rates at a PVARP of 280 and 463 +/- 29 msec were, respectively 92 +/- 5 and 66 +/- 3 msec; P < 0.0001). In the DDDR mode at a PVARP of 463 +/- 29 msec, regular ventricular pacing at 53 +/- 2 beats/min occurred due to mode switching to VVIR mode in the presence of repetitive sensed atrial events within the PVARP. One patient developed spontaneous atrial fibrillation on follow-up, which was correctly identified by the pacemaker algorithm, resulting in mode switch from DDDR to regular VVIR pacing and preservation of rate response. In conclusion, sensor controlled PVARP allows u long PVARP to be used at rest without limiting the maximum rate during exercise. In addition, to offer protection against retrograde conduction, a long PVARP and mode switching also limit the rate during atrial arrhythmias and allow regular ventricular rate responses according to the physiological demands.
引用
收藏
页码:1504 / 1514
页数:11
相关论文
共 30 条
  • [1] BAROLD SS, 1985, MODERN CARDIAC PACIN, P645
  • [2] BRUCE RA, 1971, ANN CLIN RES, V3, P323
  • [3] DYNAMIC ELECTROPHYSIOLOGY OF VENTRICULOATRIAL CONDUCTION - IMPLICATIONS FOR DDD AND DDDR PACING
    CAZEAU, S
    DAUBERT, C
    MABO, P
    RITTER, P
    LELONG, B
    POUILLOT, C
    PAILLARD, F
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (12): : 1646 - 1655
  • [4] DAVIS M, 1991, PACE PACING CLIN ELE, V14, P664
  • [5] ARRHYTHMIAS OF DUAL CHAMBER PACEMAKERS
    FURMAN, S
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1982, 5 (04): : 469 - 470
  • [6] TRACKING OF ATRIAL-FLUTTER DURING DDD PACING - ANOTHER FORM OF PACEMAKER-MEDIATED TACHYCARDIA
    GREENSPON, AJ
    GREENBERG, RM
    FRANKL, WS
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1984, 7 (06): : 955 - 960
  • [7] CLINICAL PREDICTORS AND NATURAL-HISTORY OF ATRIAL-FIBRILLATION IN PATIENTS WITH DDD PACEMAKERS
    GROSS, J
    MOSER, S
    BENEDEK, ZM
    ANDREWS, C
    FURMAN, S
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (12): : 1828 - 1831
  • [8] CHANGING EXPERIENCE WITH DUAL CHAMBER (DDD) PACEMAKERS
    HAYES, DL
    HOLMES, DR
    VLIETSTRA, RE
    OSBORN, MJ
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1984, 4 (03) : 556 - 559
  • [9] HIGANO S, 1990, RBM, V12, P35
  • [10] ILVENTO J, 1990, PACE, V13, P1199