COMPARATIVE INTRAOCULAR-PRESSURE MEASUREMENTS WITH POSITION INDEPENDENT HAND-APPLANATION TONOMETERS

被引:2
作者
KRIEGLSTEIN, GK
BRETHFELD, V
COLLANI, EV
机构
[1] UNIV WURZBURG, AUGEN KLIN, GLAUKOM ABT, D-8700 WURZBURG, FED REP GER
[2] UNIV WURZBURG, INST ANGEW MATH & STATISTIK, D-8700 WURZBURG, FED REP GER
来源
ALBRECHT VON GRAEFES ARCHIV FUR KLINISCHE UND EXPERIMENTELLE OPHTHALMOLOGIE | 1976年 / 199卷 / 02期
关键词
D O I
10.1007/BF02385206
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
A study of comparative intraocular pressure [IOP] measurements was made using the hand-held applanation tonometers of Perkins and Draeger in the sitting and lying patient. A series of 100 patients was investigated by 2 observers taking the pressures with both tonometers under blind conditions. The statistical evaluation of the data revealed an average IOP increment when lying down of 2.5-3 mm Hg for both observers and both tonometers. This postural response was fully reversible when getting up and a statistically significant overshoot-phenomenon was noted. The pressure increment when lying down was not related to initial IOP level and there was no clear relationship to the age of the patients tested. Interobserver variation and variations of both tonometers were evaluated. The effect of repeated tonometry was considered. Some aspects of position influence on IOP were discussed.
引用
收藏
页码:101 / 113
页数:13
相关论文
共 21 条
[1]  
ARMALY MF, 1963, ARCH OPHTHALMOL-CHIC, V70, P603
[2]  
BYNKE H, 1972, ACTA OPHTHALMOL, V50, P240
[3]  
Draeger J, 1966, Klin Monbl Augenheilkd, V149, P905
[4]  
DRAEGER J., 1967, INVEST OPHTHALMOM, V6, P132
[5]   PERKINS APPLANATION TONOMETER - CLINICAL AND LABORATORY EVALUATION [J].
DUNN, JS ;
BRUBAKER, RF .
ARCHIVES OF OPHTHALMOLOGY, 1973, 89 (02) :149-151
[6]  
GOLDMANN H, 1957, Ophthalmologica, V134, P221
[7]  
HETLANDE.J, 1966, ACTA OPHTHALMOL, V44, P515
[8]   INTRAOCULAR-PRESSURE AFTER CHANGING PATIENTS POSITION [J].
KINDLERLOOSLI, C ;
SCHMIDT, T .
ALBRECHT VON GRAEFES ARCHIV FUR KLINISCHE UND EXPERIMENTELLE OPHTHALMOLOGIE, 1975, 194 (01) :17-21
[9]  
KRAKAU CET, 1971, ACTA OPHTHALMOL, V49, P611
[10]   GOLDMANN APPLANATION VERSUS HAND-APPLANATION AND SCHIOTZ INDENTATION TONOMETRY [J].
KRIEGLSTEIN, GK ;
WALLER, WK .
ALBRECHT VON GRAEFES ARCHIV FUR KLINISCHE UND EXPERIMENTELLE OPHTHALMOLOGIE, 1975, 194 (01) :11-16