SELECTIVE DETERMINATION OF PENTAFLUOROBENZYL ESTER DERIVATIVES OF CARBOXYLIC-ACIDS BY GC USING MICROWAVE PLASMA AND MASS SELECTIVE DETECTION

被引:14
作者
BRILL, JH [1 ]
NARAYANAN, BA [1 ]
MCCORMICK, JP [1 ]
机构
[1] UNIV MISSOURI,DEPT CHEM,COLUMBIA,MO 65211
关键词
MICROWAVE-INDUCED PLASMA DETECTOR; FLUORINE; CARBOXYLIC ACIDS; PENTAFLUOROBENZYL ESTERS;
D O I
10.1366/0003702914335193
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A highly selective, sensitive method for determination of carboxylic acids in water has been developed that is based on fluorine-selective gas chromatography (GC) detection of pentafluorobenzyl ester derivatives. The atmospheric-pressure helium microwave-induced plasma detector is well suited for fluorine-specific detection of these derivatives, which were well resolved by capillary GC and displayed a characteristic fragmentation behavior when examined by GC-mass selective detection. For demonstration purposes, a number of carboxylic acids, ranging from C1 to C20, were examined and found to be derivatized with yields greater than 90%, and a sample of creek water was examined for its carboxylic acid content.