WEAR OF THE POLYETHYLENE ACETABULAR CUP - METALLIC AND CERAMIC HEADS COMPARED IN A HIP SIMULATOR

被引:87
作者
SAIKKO, VO [1 ]
PAAVOLAINEN, PO [1 ]
SLATIS, P [1 ]
机构
[1] ORTHOPED HOSP INVALID FDN,HELSINKI,FINLAND
来源
ACTA ORTHOPAEDICA SCANDINAVICA | 1993年 / 64卷 / 04期
基金
芬兰科学院;
关键词
D O I
10.3109/17453679308993653
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Ultra-high molecular weight polyethylene acetabular cups of 5 different total hip systems (Muller, Mallory-Head, Lubinus, P.C.A. and Charnley-Elite) were worn on a new 5-station hip joint simulator. The cups articulated against modular metallic (stainless steel in Muller and Charnley-Elite, ion-implanted Ti-6Al-4V in Mallory-Head, and Co-Cr-Mo in Lubinus and P.C.A.) and modular alumina ceramic femoral heads for 3 million walking cycles. The mean wear rate of cups against alumina heads (range 0-5.7 mg/10(6) cycles, corresponding to 0-0.008 mm/year) was usually lower than against metallic heads (range 3.9-178 mg/10(6) cycles, corresponding to 0.005-0.24 mm/year). In the metal-head prostheses, the mean wear rate was highest against stainless steel heads, and lowest against ion-implanted Ti-6Al-4V heads. As the wear rates are compared with published clinical observations, it can be concluded that the hip joint simulator is capable of producing realistic wear rates; it is a useful instrument in the study of the wear behavior of new designs, materials, surface treatments and coatings prior to clinical trials. However, the taper-fit attachment of modular heads proved problematical, showing corrosion and wear at the conical head-spigot interface.
引用
收藏
页码:391 / 402
页数:12
相关论文
共 30 条
[1]  
[Anonymous], 1992, SOCIOLOGICA, V7
[2]   LABORATORY WEAR TESTS AND CLINICAL OBSERVATIONS OF THE PENETRATION OF FEMORAL HEADS INTO ACETABULAR CUPS IN TOTAL REPLACEMENT HIP JOINTS .3. THE MEASUREMENT OF INTERNAL VOLUME CHANGES IN EXPLANTED CHARNLEY SOCKETS AFTER 2-16 YEARS INVIVO AND THE DETERMINATION OF WEAR FACTORS [J].
ATKINSON, JR ;
DOWSON, D ;
ISAAC, JH ;
WROBLEWSKI, BM .
WEAR, 1985, 104 (03) :225-244
[3]  
CLARKE IC, 1982, CRIT REV BIOMED ENG, V8, P29
[4]   FLUID-SORPTION PHENOMENA IN STERILIZED POLYETHYLENE ACETABULAR PROSTHESES [J].
CLARKE, IC ;
STARKEBAUM, W ;
HOSSEINIAN, A ;
MCGUIRE, P ;
OKUDA, R ;
SALOVEY, R ;
YOUNG, R .
BIOMATERIALS, 1985, 6 (03) :184-188
[5]  
CLARKE IC, 1983, ASTM STP, V796, P136
[6]  
COLLIER JP, 1991, CLIN ORTHOP RELAT R, V271, P305
[7]  
DAVIDSON JA, 1989, 8TH SO BIOM ENG C RI
[8]  
DAVIDSON JA, 1992, JAP J TRIBOLOGY, V37, P399
[9]   CHARACTERISTICS OF ACETABULAR CUPS WORN IN HUMAN-BODY [J].
DOWLING, JM ;
ATKINSON, JR ;
DOWSON, D ;
CHARNLEY, J .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1978, 60 (03) :375-382
[10]  
ISAAC GH, 1992, CLIN ORTHOP RELAT R, V276, P115