INVIVO BEHAVIOR OF HYPODERMICALLY IMPLANTED MICROFABRICATED GLUCOSE SENSORS

被引:70
作者
KOUDELKA, M
ROHNERJEANRENAUD, F
TERRETTAZ, J
BOBBIONIHARSCH, E
DEROOIJ, NF
JEANRENAUD, B
机构
[1] UNIV GENEVA,METAB RES LABS,CH-1211 GENEVA 4,SWITZERLAND
[2] FAC MED GENEVA,METAB RES LABS,GENEVA,SWITZERLAND
关键词
GLUCOSE SENSOR; AMPEROMETRIC ENZYME ELECTRODE; SUBCUTANEOUS GLUCOSE; INVIVO EXPERIMENTS;
D O I
10.1016/0956-5663(91)85005-H
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The in-vivo behaviour of microfabricated GOD (glucose oxidase)/H2O2 glucose sensor implanted subcutaneously in normal anaesthetized rats has been studied. The sensor consists of a planar, three-electrode microcell, an enzyme membrane (glucose oxidase and bovine serum albumin cross-linked with glutaraldehyde) and an outer diffusion limiting polyurethane membrane. The sensor behaviour during hyperglycaemic (13.8 mM and 11.2 mM), euglycaemic (7.8 mM) and hypoglycaemic (3.5 mM) plateau levels was determined. The values of the in-vivo sensitivity (0.64 +/- 0.05 nA/mM) and background current (1.25 +/- 0.4 nA) were determined using a two-point calibration method and then used to calculate apparent subcutaneous glucose concentrations. The results show the presence of a good correlation between all the plasma glucose levels (G) and the apparent subcutaneous tissue concentrations (G'), with G' = 0.997. G - 0.066, r = 0.9782.
引用
收藏
页码:31 / 36
页数:6
相关论文
共 14 条
[1]   SUBCUTANEOUS IMPLANTATION OF A FERROCENE-MEDIATED GLUCOSE SENSOR IN PIGS [J].
CLAREMONT, DJ ;
SAMBROOK, IE ;
PENTON, C ;
PICKUP, JC .
DIABETOLOGIA, 1986, 29 (11) :817-821
[2]   PHYSIOLOGICAL PREPARATION FOR STUDYING THE RESPONSE OF SUBCUTANEOUSLY IMPLANTED GLUCOSE AND OXYGEN SENSORS [J].
ERTEFAI, S ;
GOUGH, DA .
JOURNAL OF BIOMEDICAL ENGINEERING, 1989, 11 (05) :362-368
[3]   ASSESSMENT OF SUBCUTANEOUS GLUCOSE-CONCENTRATION - VALIDATION OF THE WICK TECHNIQUE AS A REFERENCE FOR IMPLANTED ELECTROCHEMICAL SENSORS IN NORMAL AND DIABETIC DOGS [J].
FISCHER, U ;
ERTLE, R ;
ABEL, P ;
REBRIN, K ;
BRUNSTEIN, E ;
VONDORSCHE, HH ;
FREYSE, EJ .
DIABETOLOGIA, 1987, 30 (12) :940-945
[4]   FABRICATION AND CHARACTERIZATION OF A PLANAR ELECTROCHEMICAL-CELL AND ITS APPLICATION AS A GLUCOSE SENSOR [J].
GERNET, S ;
KOUDELKA, M ;
DEROOIJ, NF .
SENSORS AND ACTUATORS, 1989, 18 (01) :59-70
[5]  
ITO K, 1986, ACS SYM SER, V309, P373
[6]   CHARACTERIZATION BY MICRODIALYSIS OF INTERCELLULAR GLUCOSE LEVEL IN SUBCUTANEOUS TISSUE IN HUMANS [J].
JANSSON, PA ;
FOWELIN, J ;
SMITH, U ;
LONNROTH, P .
AMERICAN JOURNAL OF PHYSIOLOGY, 1988, 255 (02) :E218-E220
[7]   PLANAR AMPEROMETRIC ENZYME-BASED GLUCOSE MICROELECTRODE [J].
KOUDELKA, M ;
GERNET, S ;
DEROOIJ, NF .
SENSORS AND ACTUATORS, 1989, 18 (02) :157-165
[8]   AN AMPEROMETRIC NEEDLE-TYPE GLUCOSE SENSOR TESTED IN RATS AND MAN [J].
MATTHEWS, DR ;
BOWN, E ;
BECK, TW ;
PLOTKIN, E ;
LOCK, L ;
GOSDEN, E ;
WICKHAM, M .
DIABETIC MEDICINE, 1988, 5 (03) :248-252
[9]  
PFEIFFER EF, 1988, HORMONE METABOL RE S, V20
[10]   INVIVO MOLECULAR SENSING IN DIABETES-MELLITUS - AN IMPLANTABLE GLUCOSE SENSOR WITH DIRECT ELECTRON-TRANSFER [J].
PICKUP, JC ;
SHAW, GW ;
CLAREMONT, DJ .
DIABETOLOGIA, 1989, 32 (03) :213-217