LOGARITHMIC AMPLIFIERS

被引:22
作者
GANDLER, W [1 ]
SHAPIRO, H [1 ]
机构
[1] NIH,DIV COMP RES & TECHNOL,COMP SYST LAB,W NEWTON,MA
来源
CYTOMETRY | 1990年 / 11卷 / 03期
关键词
calibration; coefficient of variation; Dynamic range;
D O I
10.1002/cyto.990110317
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Logarithmic amplifiers (log amps), which produce an output signal proportional to the logarithm of the input signal, are widely used in cytometry for measurements of parameters that vary over a wide dynamic range, e.g., cell surface immunofluorescence. Existing log amp circuits all deviate to some extent from ideal performance with respect to dynamic range and fidelity to the logarithmic curve; accuracy in quantitative analysis using log amps therefore requires that log amps be individually calibrated. However, accuracy and precision may be limited by photon statistic and system noise when very low level input signals are encountered. Copyright © 1990 Wiley‐Liss, Inc.
引用
收藏
页码:447 / 450
页数:4