ENERGY RESOLUTION CALCULATION OF THE PWO CALORIMETER, COMPARISON WITH THE BEAM TESTS

被引:7
作者
PROKOSHKIN, YD [1 ]
SHTANNIKOV, AV [1 ]
机构
[1] INST HIGH ENERGY PHYS, PROTVINO, RUSSIA
关键词
D O I
10.1016/0168-9002(95)00226-X
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The basic parameters of a PbWO4 crystal calorimeter are GEANT evaluated to optimize its energy resolution in the photon energy range from tens of MeV to hundreds of GeV. The energy resolution is well factorized in this region as a function of scintillation efficiency, thanks to a high transparency of the crystal. The calculated energy dependence of the resolution fits the results of the first beam measurements.