CAVERNOUS HEMANGIOMAS - DIPOLAR SUSCEPTIBILITY ARTIFACTS AT MR-IMAGING

被引:36
作者
KIM, JK [1 ]
KUCHARCZYK, W [1 ]
HENKELMAN, RM [1 ]
机构
[1] UNIV TORONTO,SUNNYBROOK HLTH SCI CTR,DEPT RADIOL,TORONTO M4N 3M5,ON,CANADA
关键词
ANGIOMA; CENTRAL NERVOUS SYSTEM; MAGNETIC RESONANCE (MR); ARTIFACT; EXPERIMENTAL; PHYSICS;
D O I
10.1148/radiology.187.3.8497623
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A misleading susceptibility artifact was originally encountered at magnetic resonance imaging in a patient with multiple cavernous hemangiomas. By using a gradient-recalled echo sequence, the artifact appeared on axial sections as a single ring of enhanced signal intensity within the expected signal void around many of the individual lesions. It was hypothesized and confirmed through phantom experiments with ferromagnetic samples that hemosiderin deposits within each lesion behaved like a point magnetic dipole, causing the intravoxel signal interference patterns that appeared as a ring of enhanced signal intensity. The artifact was therefore found to be entirely susceptibility-induced. Knowledge of the origin and appearance of this new artifact will aid radiologists in the interpretation of susceptibility images and help them avoid misinterpreting a single lesion as multiple contiguous lesions.
引用
收藏
页码:735 / 741
页数:7
相关论文
共 10 条
[1]   NUMERICAL-ANALYSIS OF THE MAGNETIC-FIELD FOR ARBITRARY MAGNETIC-SUSCEPTIBILITY DISTRIBUTIONS IN 2D [J].
BHAGWANDIEN, R ;
VANEE, R ;
BEERSMA, R ;
BAKKER, CJG ;
MOERLAND, MA ;
LAGENDIJK, JJW .
MAGNETIC RESONANCE IMAGING, 1992, 10 (02) :299-313
[2]  
GHOSH P, 1992, SOC MAGNETIC RESONAN, P344
[3]   OCCULT CEREBRAL VASCULAR MALFORMATIONS - HIGH-FIELD MR IMAGING [J].
GOMORI, JM ;
GROSSMAN, RI ;
GOLDBERG, HI ;
HACKNEY, DB ;
ZIMMERMAN, RA ;
BILANIUK, LT .
RADIOLOGY, 1986, 158 (03) :707-713
[4]   CAUSE OF SIGNAL LOSS IN MR IMAGES OF OLD HEMORRHAGIC LESIONS [J].
HARDY, PA ;
KUCHARCZYK, W ;
HENKELMAN, RM .
RADIOLOGY, 1990, 174 (02) :549-555
[5]  
Ludeke K M, 1985, Magn Reson Imaging, V3, P329, DOI 10.1016/0730-725X(85)90397-2
[6]  
PLONUS MA, 1978, APPL ELECTROMAGNETIC, P297
[7]   DIRECT IMAGING OF MAGNETIC-FIELD GRADIENTS BY GROUP SPIN-ECHO SELECTION [J].
POSSE, S .
MAGNETIC RESONANCE IN MEDICINE, 1992, 25 (01) :12-29
[8]   SUSCEPTIBILITY ARTIFACTS IN SPIN-ECHO AND GRADIENT-ECHO IMAGING [J].
POSSE, S ;
AUE, WP .
JOURNAL OF MAGNETIC RESONANCE, 1990, 88 (03) :473-492
[9]   CEREBRAL CAVERNOUS MALFORMATIONS - INCIDENCE AND FAMILIAL OCCURRENCE [J].
RIGAMONTI, D ;
HADLEY, MN ;
DRAYER, BP ;
JOHNSON, PC ;
HOENIGRIGAMONTI, K ;
KNIGHT, JT ;
SPETZLER, RF .
NEW ENGLAND JOURNAL OF MEDICINE, 1988, 319 (06) :343-347
[10]   MR IMAGING OF SUSCEPTIBILITY-INDUCED MAGNETIC-FIELD INHOMOGENEITIES [J].
WENDT, RE ;
WILCOTT, MR ;
NITZ, W ;
MURPHY, PH ;
BRYAN, RN .
RADIOLOGY, 1988, 168 (03) :837-841