SIMULTANEOUS FLOW-INJECTION DETERMINATION OF NITRATE AND NITRITE IN WATER BY GAS-PHASE CHEMILUMINESCENCE

被引:40
作者
AOKI, T [1 ]
WAKABAYASHI, M [1 ]
机构
[1] KYOTO UNIV,DEPT SCI,FUSHIMI,KYOTO 612,JAPAN
关键词
CHEMILUMINESCENCE; FLOW INJECTION; NITRATE; NITRITE; WATERS;
D O I
10.1016/0003-2670(94)00603-J
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A reagent-injection now injection procedure for the determination of nitrate and nitrite in water by reduction to NO followed by a membrane separation process and chemiluminescence detection, with air-stripping, is proposed. Microporous PTFE was used as a membrane to transfer NO to the gas phase after reducing nitrate and nitrite with Ti(III) and nitrite alone by iodide in acidic media. Chemiluminescence (CL) peak height signals produced by the reaction of NO with ozone were linearly proportional to concentrations from 5 mu g N/l as nitrate and 3 mu g N/l as nitrite to around 1 mg N/l. The relative standard deviation (n = 5) was 1.2% at 0.70 mg N/l nitrate and 1.4% at 0.28 mg N/l nitrite. The response time from injecting the reducing agent until the signal reached a peak was 1.0 min. In the presence of nitrite, nitrate could be determined from the difference between the CL intensity in the Ti(III) and iodide systems.
引用
收藏
页码:308 / 312
页数:5
相关论文
共 15 条
[2]  
[Anonymous], 1972, B FISH RES BOARD CAN
[3]   CONTINUOUS-FLOW FLUOROMETRIC-DETERMINATION OF AMMONIA IN WATER [J].
AOKI, T ;
UEMURA, S ;
MUNEMORI, M .
ANALYTICAL CHEMISTRY, 1983, 55 (09) :1620-1622
[4]   CONTINUOUS-FLOW DETERMINATION OF NITRITE WITH MEMBRANE SEPARATION CHEMILUMINESCENCE DETECTION [J].
AOKI, T .
BIOMEDICAL CHROMATOGRAPHY, 1990, 4 (03) :128-130
[6]   CONTINUOUS-FLOW DETERMINATION OF FREE CHLORINE IN WATER [J].
AOKI, T ;
MUNEMORI, M .
ANALYTICAL CHEMISTRY, 1983, 55 (02) :209-212
[7]   CONTINUOUS-FLOW METHOD FOR THE DETERMINATION OF TOTAL INORGANIC CARBONATE IN WATER [J].
AOKI, T ;
FUJIMARU, Y ;
OKA, Y ;
FUJIE, K .
ANALYTICA CHIMICA ACTA, 1993, 284 (01) :167-171
[8]   MERGING ZONES IN FLOW INJECTION ANALYSIS .1. DOUBLE PROPORTIONAL INJECTOR AND REAGENT CONSUMPTION [J].
BERGAMIN, H ;
ZAGATTO, EAG ;
KRUG, FJ ;
REIS, BF .
ANALYTICA CHIMICA ACTA, 1978, 101 (01) :17-23
[9]   CONFIGURATION WITH INTERNALLY COUPLED VALVES TO OVERCOME SHORTCOMINGS IN THE SIMULTANEOUS DETERMINATION OF NITRITE AND NITRATE BY FLOW-INJECTION ANALYSIS [J].
BERMUDEZ, B ;
RIOS, A ;
DECASTRO, MDL ;
VALCARCEL, M .
TALANTA, 1988, 35 (10) :810-812