ANALYSIS OF T2 LIMITATIONS AND OFF-RESONANCE EFFECTS ON SPATIAL-RESOLUTION AND ARTIFACTS IN ECHO-PLANAR IMAGING

被引:294
作者
FARZANEH, F [1 ]
RIEDERER, SJ [1 ]
PELC, NJ [1 ]
机构
[1] GE,MED SYST,MILWAUKEE,WI 53201
关键词
D O I
10.1002/mrm.1910140112
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Several aspects of blipped echo‐planar imaging (EPI) are treated mathematically. An expression relating the necessary readout gradient strength and sampling time to the spatial resolution and readout duration is derived. It is shown how the net spatial resolution may be limited by the object's T2 characteristics and Bo field homogeneity, irrespective of the number of sampled points. Additionally, off‐resonance effects result in a loss of spatial resolution and image distortion to a considerably greater degree than in conventional two‐dimensional Fourier transform imaging. The extent of these effects is directly related to the time required to acquire the data matrix, and is therefore amplified when EPI is implemented on a standard commercial whole‐body system which because of limited gradient performance uses necessarily longer sampling durations. Specific hardware modifications to a standard commercial imager are considered to allow successful EPI implementation. EPI image characteristics are compared quantitatively with those of conventional methods. © 1990 Academic Press, Inc. Copyright © 1990 Wiley‐Liss, Inc., A Wiley Company
引用
收藏
页码:123 / 139
页数:17
相关论文
共 15 条
[1]   HIGH-SPEED SPIRAL-SCAN ECHO PLANAR NMR IMAGING .1. [J].
AHN, CB ;
KIM, JH ;
CHO, ZH .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1986, 5 (01) :2-7
[2]   REAL-TIME MOVIE IMAGING FROM A SINGLE CARDIAC CYCLE BY NMR [J].
CHAPMAN, B ;
TURNER, R ;
ORDIDGE, RJ ;
DOYLE, M ;
CAWLEY, M ;
COXON, R ;
GLOVER, P ;
MANSFIELD, P .
MAGNETIC RESONANCE IN MEDICINE, 1987, 5 (03) :246-254
[3]   INFLUENCE OF PHYSIOLOGICAL MOTION ON THE APPEARANCE OF TISSUE IN MR IMAGES [J].
EHMAN, RL ;
MCNAMARA, MT ;
BRASCH, RC ;
FELMLEE, JP ;
GRAY, JE ;
HIGGINS, CB .
RADIOLOGY, 1986, 159 (03) :777-782
[4]   FLASH IMAGING - RAPID NMR IMAGING USING LOW FLIP-ANGLE PULSES [J].
HAASE, A ;
FRAHM, J ;
MATTHAEI, D ;
HANICKE, W ;
MERBOLDT, KD .
JOURNAL OF MAGNETIC RESONANCE, 1986, 67 (02) :258-266
[5]   THE LIMITATIONS OF NMR RECALLED-ECHO IMAGING TECHNIQUES [J].
JOHNSON, G ;
HUTCHISON, JMS .
JOURNAL OF MAGNETIC RESONANCE, 1985, 63 (01) :14-30
[6]   MULTI-PLANAR IMAGE-FORMATION USING NMR SPIN ECHOES [J].
MANSFIELD, P .
JOURNAL OF PHYSICS C-SOLID STATE PHYSICS, 1977, 10 (03) :L55-L58
[7]  
MANSFIELD P, 1988, 7TH SOC MAGN RES MED, P237
[8]   REAL-TIME MOVIE IMAGES BY NMR [J].
ORDIDGE, RJ ;
MANSFIELD, P ;
DOYLE, M ;
COUPLAND, RE .
BRITISH JOURNAL OF RADIOLOGY, 1982, 55 (658) :729-733
[9]  
PYKETT IL, 1987, MAGNET RESON MED, V5, P563
[10]  
RZEDZIAN R, 1983, LANCET, V2, P1281