ARTIFACTS IN INTRAVASCULAR ULTRASOUND IMAGING - ANALYSES AND IMPLICATIONS

被引:39
作者
FINET, G
MAURINCOMME, E
TABIB, A
CROWLEY, RJ
MAGNIN, I
RORIZ, R
BEAUNE, J
AMIEL, M
机构
[1] HOSP CIVILS LYON,HOP CARDIOVASC & PNEUMOL,DEPT HEMODYNAM,LYON,FRANCE
[2] HOSP CIVILS LYON,HOP CARDIOVASC & PNEUMOL,DEPT RADIOL,LYON,FRANCE
[3] CNRS,UNIT 1216,F-69373 LYON,FRANCE
[4] BOSTON SCI CORP,DIV ULTRASOUND,WATERTOWN,MA 02172
关键词
ULTRASOUND; INTRAVASCULAR ULTRASOUND; ECHOCARDIOGRAPHY; ARTIFACTS; IMAGE PROCESSING; IN-VITRO VALIDATION; ULTRASONIC TISSUE EFFECTS;
D O I
10.1016/0301-5629(93)90077-2
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The ability of an intravascular ultrasound catheter to give cross-sectional images of vessel walls and surrounding tissues, and the behavior of ultrasound in heterogeneous media, are at the origin of degradation of image quality. Qualitative and quantitative analyses of in vivo studies are then operator-dependent and are limited by artifacts. We investigated these limitations by an in vitro study on plexiglass phantoms and segments of fresh arteries. We used a 20 MHz transducer mounted on the tip of a 4.8 F catheter and an interventional ultrasound system. The ultrasound beam is reflected onto the rotating transducer at 600 rotations per minute (RPM), creating 360-degrees real-time images (10 images/second). We then observed, analyzed and interpreted the most specific reasons for image artifacts: geometric distortions, multiple echoes, the point spread function (PSF) of the imaging system, near-field effects, ''petal-shaped'' effect, and ultrasound speckle. Various practical implications have resulted from this study. Only a thorough knowledge of how to avoid some of the most obvious pitfalls will enable the user to obtain maximum benefits from intravascular ultrasound imaging, and to appreciate its limitations.
引用
收藏
页码:533 / 547
页数:15
相关论文
共 35 条
  • [1] ULTRASONIC INTRACARDIAC SCANNER
    BOM, N
    LANCEE, CT
    VANEGMON.FC
    [J]. ULTRASONICS, 1972, 10 (02) : 72 - &
  • [2] CHAE J-S, 1990, Journal of the American College of Cardiology, V15, p28A
  • [3] Crowley R J, 1989, Int J Card Imaging, V4, P145, DOI 10.1007/BF01745144
  • [4] Crowley R J, 1991, Int J Card Imaging, V6, P145, DOI 10.1007/BF01797847
  • [5] ANGLE-DEPENDENT BACKSCATTER FROM THE ARTERIAL-WALL
    DEKROON, MGM
    VANDERWAL, LF
    GUSSENHOVEN, WJ
    BOM, N
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1991, 17 (02) : 121 - 126
  • [6] FITZGERALD PJ, 1991, J AM COLL CARDIOL S, V17, pA217
  • [7] FITZGERALD PJ, 1991, CIRCULATION, V4, pA438
  • [8] ARTERIAL-WALL CHARACTERISTICS DETERMINED BY INTRAVASCULAR ULTRASOUND IMAGING - AN INVITRO STUDY
    GUSSENHOVEN, EJ
    ESSED, CE
    LANCEE, CT
    MASTIK, F
    FRIETMAN, P
    VANEGMOND, FC
    REIBER, J
    BOSCH, H
    VANURK, H
    ROELANDT, J
    BOM, N
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (04) : 947 - 952
  • [9] Gussenhoven W J, 1989, Int J Card Imaging, V4, P105, DOI 10.1007/BF01745140
  • [10] WHAT MIGHT ECHOGRAPHY LEARN FROM IMAGE SCIENCE
    HILL, CR
    BAMBER, JC
    CRAWFORD, DC
    LOWE, HJ
    WEBB, S
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1991, 17 (06) : 559 - 575