BASIC STUDY TO DEVELOP THE UNDULATION PUMP FOR PRACTICAL USE - ANTITHROMBOGENICITY, HEMOLYSIS, AND FLOW PATTERNS INSIDE THE PUMP

被引:12
作者
ABE, Y
CHINZEI, T
ISOYAMA, T
ONE, T
MABUCHI, K
IMANISHI, K
KOUNO, A
ATSUMI, K
FUJIMASA, I
IMACHI, K
机构
[1] UNIV TOKYO,ADV SCI & TECHNOL RES CTR,TOKYO,JAPAN
[2] TOKYO WOMENS MED COLL,HOSP 2,TOKYO 162,JAPAN
关键词
UNDULATION PUMP; PRECESSIONAL DISPLACEMENT PUMP; HEMOLYSIS; THROMBUS FORMATION; FLOW VISUALIZATION STUDY;
D O I
10.1111/j.1525-1594.1995.tb02404.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
The undulation pump (formerly called the precessional displacement pump) is a continuous flow displacement-type blood pump that is being developed as an implantable total artificial heart. A new undulation pump was developed for chronic use and was examined with animal experiment and flow visualization studies. In the animal experiment using a left ventricular bypass in goats, severe hemolysis occurred. After driving for 12 h, thrombus formation inside the pump was found. In the flow visualization studies, the flow pattern showed that the flow inside the pump was a very complicated turbulent flow. Improvement of hemolysis and thrombus formation is important to realize implantable total artificial hearts using undulation pumps.
引用
收藏
页码:691 / 693
页数:3
相关论文
共 3 条
[1]
Abe Y., 1993, Japanese Journal of Artificial Organs, V22, P683
[2]
Abe Y., 1994, Japanese Journal of Artificial Organs, V23, P25
[3]