CHARACTERIZATION OF POTASSIUM-SELECTIVE OPTODE MEMBRANES BASED ON NEUTRAL IONOPHORES AND APPLICATION IN HUMAN BLOOD-PLASMA

被引:85
作者
WANG, K [1 ]
SEILER, K [1 ]
MORF, WE [1 ]
SPICHIGER, UE [1 ]
SIMON, W [1 ]
LINDNER, E [1 ]
PUNGOR, E [1 ]
机构
[1] TECH UNIV BUDAPEST,INST GEN & ANALYT CHEM,H-1111 BUDAPEST 11,HUNGARY
关键词
Chromoionophores; ion-selective membranes; optode membranes; potassium concentration in plasma;
D O I
10.2116/analsci.6.715
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Plasticized PVC membranes incorporating conventional potassium-selective neutral ionophores and a neutral H+-selective chromoionophore have been used as reversible sensing devices for an optical determination of K+ activities in pH-buffered solutions. Dynamic range, reproducibility, response time, and selectivity of the new optode membranes are discussed. With this potassium-selective membrane, the reversible optical determination of total K+ concentration in human blood plasma is available. Results are compared to measurements with ion selective electrodes and flame photometry. © 1990, The Japan Society for Analytical Chemistry. All rights reserved.
引用
收藏
页码:715 / 720
页数:6
相关论文
共 20 条
  • [1] AN OPTICAL POTASSIUM-ION SENSOR
    ALDER, JF
    ASHWORTH, DC
    NARAYANASWAMY, R
    MOSS, RE
    SUTHERLAND, IO
    [J]. ANALYST, 1987, 112 (08) : 1191 - 1192
  • [2] AMMANN D, 1983, ION SEL ELECTRODE R, V5, P3
  • [3] BERRY MN, 1989, CLIN CHEM, V35, P817
  • [4] CHARLTON SC, 1982, CLIN CHEM, V28, P1857
  • [5] CHEMICAL SENSORS FOR BEDSIDE MONITORING OF CRITICALLY ILL PATIENTS
    COLLISON, ME
    MEYERHOFF, ME
    [J]. ANALYTICAL CHEMISTRY, 1990, 62 (07) : A425 - &
  • [6] DINTEN O, UNPUB
  • [7] NEW CHROMOIONOPHORES
    DIX, JP
    VOGTLE, F
    [J]. CHEMISCHE BERICHTE-RECUEIL, 1981, 114 (02): : 638 - 651
  • [8] HIDEO S, 1977, TALANTA, V24, P1763
  • [9] KUMAR A, 1988, CLIN CHEM, V34, P1709
  • [10] Massart D. L., 1988, CHEMOMETRICS TXB