A SCREENING PROTOCOL FOR THE DIRECT DETERMINATION OF LOW PPB LEVELS OF URANYL CATION USING ARSENAZO-III AND CAPILLARY ELECTROPHORESIS

被引:19
作者
COLBURN, BA [1 ]
SEPANIAK, MJ [1 ]
HINTON, ER [1 ]
机构
[1] LOCKHEED MARTIN ENERGY SYST,ANALYT SERV ORG,OAK RIDGE,TN 37831
来源
JOURNAL OF LIQUID CHROMATOGRAPHY | 1995年 / 18卷 / 18-19期
关键词
D O I
10.1080/10826079508014620
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Capillary electrophoresis is used as a screening protocol for the determination of uranyl ion complexed with Arsenate III. Long sample plugs are injected and stacked to narrow zones. Capillary column inner diameter, running buffer constituents and concentrations, effective column length and injected sample plug length are optimized and discussed. A limit of detection of 10 ppb is achieved with a linear dynamic range of nearly 3 orders of magnitude in concentration injected. Analytical figures of merit are presented and river water samples are analyzed.
引用
收藏
页码:3699 / 3718
页数:20
相关论文
共 30 条
[1]   DETERMINATION OF URANIUM AND THORIUM CONCENTRATIONS IN SOILS - COMPARISON OF ISOTOPE-DILUTION SECONDARY ION MASS-SPECTROMETRY AND ISOTOPE-DILUTION THERMAL IONIZATION MASS-SPECTROMETRY [J].
ADRIAENS, AG ;
FASSETT, JD ;
KELLY, WR ;
SIMONS, DS ;
ADAMS, FC .
ANALYTICAL CHEMISTRY, 1992, 64 (23) :2945-2950
[2]   SENSITIVITY ENHANCEMENT FOR CAPILLARY ELECTROPHORESIS [J].
ALBIN, M ;
GROSSMAN, PD ;
MORING, SE .
ANALYTICAL CHEMISTRY, 1993, 65 (10) :A489-A497
[3]  
BALTISBERGER RJ, 1964, ANAL CHEM, V36, P2369
[4]   SPECTROPHOTOMETRIC DETERMINATION OF URANIUM IN NATURAL-WATERS [J].
BERMEJOBARRERA, A ;
YEBRABIURRUN, MC ;
FRAGATRILLO, LM .
ANALYTICA CHIMICA ACTA, 1990, 239 (02) :321-323
[5]   DIRECT DETECTION OF TRACE LEVELS OF URANIUM BY LASER-INDUCED KINETIC PHOSPHORIMETRY [J].
BRINA, R ;
MILLER, AG .
ANALYTICAL CHEMISTRY, 1992, 64 (13) :1413-1418
[6]   LARGE-VOLUME STACKING OF ANIONS IN CAPILLARY ELECTROPHORESIS USING AN ELECTROOSMOTIC FLOW MODIFIER AS A PUMP [J].
BURGI, DS .
ANALYTICAL CHEMISTRY, 1993, 65 (24) :3726-3729
[7]   OPTIMIZATION IN SAMPLE STACKING FOR HIGH-PERFORMANCE CAPILLARY ELECTROPHORESIS [J].
BURGI, DS ;
CHIEN, RL .
ANALYTICAL CHEMISTRY, 1991, 63 (18) :2042-2047
[8]   SAMPLE STACKING OF AN EXTREMELY LARGE INJECTION VOLUME IN HIGH-PERFORMANCE CAPILLARY ELECTROPHORESIS [J].
CHIEN, RL ;
BURGI, DS .
ANALYTICAL CHEMISTRY, 1992, 64 (09) :1046-1050
[9]   ON-COLUMN SAMPLE CONCENTRATION USING FIELD AMPLIFICATION IN CZE [J].
CHIEN, RL ;
BURGI, DS .
ANALYTICAL CHEMISTRY, 1992, 64 (08) :A489-A496
[10]   SEPARATIONS OF DERIVATIZED AMINO-ACID ENANTIOMERS BY CYCLODEXTRIN-MODIFIED CAPILLARY ELECTROPHORESIS - MECHANISTIC AND MOLECULAR MODELING STUDIES [J].
COPPER, CL ;
DAVIS, JB ;
COLE, RO ;
SEPANIAK, MJ .