NEW GIANTURCO-GRIFKA VASCULAR OCCLUSION DEVICE - INITIAL STUDIES IN A CANINE MODEL

被引:44
作者
GRIFKA, RG
MULLINS, CE
GIANTURCO, C
NIHILL, MR
OLAUGHLIN, MP
SLACK, MC
CLUBB, FJ
MYERS, TJ
机构
[1] TEXAS HEART INST,HOUSTON,TX 77030
[2] TEXAS CHILDRENS HOSP,BAYLOR COLL MED,LILLIE FRANK ABERCROMBIE SECT PEDIAT CARDIOL,HOUSTON,TX 77030
关键词
HEART DEFECTS; CONGENITAL; CATHETERIZATION; OCCLUSION; DUCTUS ARTERIOSUS; PATENT;
D O I
10.1161/01.CIR.91.6.1840
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Transcatheter closure of cardiovascular defects remains a challenge. Several occlusion devices are available, but each device has limitations. The purpose of this study was to evaluate the new Gianturco-Grifka vascular occlusion device (GGVOD) in a canine model. Methods and Results A total of 26 GGVODs were implanted as part of short- and long-term studies. In the shortterm study, 1 GGVOD was implanted in each of 11 systemic arteries from 3.2 to 9.0 mm in diameter. All 11 arteries were occluded immediately. In the long-term study, an aortopulmonary shunt was placed in 10 dogs (9, Gore-tex graft; 1, subclavian artery) followed by GGVOD implantation; additionally, a GGVOD was implanted in 5 subclavian arteries. The dogs were bearded for 3 to 6 months, then recatheterized and euthanatized. Immediately after implantation, the 5 subclavian arteries and 9 Gore-tex shunts were occluded completely; the 1 subclavian artery shunt had a small residual leak. At recatheterization, all 10 shunts and 5 subclavian arteries were occluded completely. Necropsy revealed all shunts to be occluded, with the aortic and pulmonic orifices covered with a neointimal layer. The mean fluoroscopic time needed for GGVOD implantation was 9 minutes (range, 3 to 22 minutes). Conclusions (1) In a canine model, the GGVOD is effective for transcatheter occlusion of arteries and aortopulmonary shunts from 3 to 9 mm in diameter. Possible indications in children include aortopulmonary collateral vessels, long patent ductus arteriosus, systemic-pulmonary shunts, AV malformations, and arteries supplying tumors. (2) GGVOD implantation requires a short fluoroscopic time.
引用
收藏
页码:1840 / 1846
页数:7
相关论文
共 19 条
[1]   MINI GIANTURCO STAINLESS-STEEL COILS FOR TRANSCATHETER VASCULAR OCCLUSION [J].
ANDERSON, JH ;
WALLACE, S ;
GIANTURCO, C ;
GERSON, LP .
RADIOLOGY, 1979, 132 (02) :301-303
[2]   EMBOLOTHERAPY OF PULMONARY ARTERIOVENOUS-MALFORMATIONS WITH DETACHABLE BALLOONS [J].
BARTH, KH ;
WHITE, RI ;
KAUFMAN, SL ;
TERRY, PB ;
ROLAND, JM .
RADIOLOGY, 1982, 142 (03) :599-606
[3]  
BASH SE, 1984, CIRCULATION, V70, P285
[4]  
CAREY LS, 1974, J CAN ASSOC RADIOL, V25, P113
[5]  
CHUANG VP, 1975, RADIOLOGY, V115, P55
[6]  
GRIFKA RG, 1992, J AM COLL CARDIOL, V7, pA117
[7]  
KING TD, 1974, SURGERY, V75, P383
[8]  
LOCK JE, 1987, CIRCULATION, V76, P28
[9]   TRANSCATHETER UMBRELLA CLOSURE OF CONGENITAL HEART-DEFECTS [J].
LOCK, JE ;
COCKERHAM, JT ;
KEANE, JF ;
FINLEY, JP ;
WAKELY, PE ;
FELLOWS, KE .
CIRCULATION, 1987, 75 (03) :593-599
[10]   A NEW DEVICE FOR TRANSCATHETER CLOSURE OF THE PATENT DUCTUS-ARTERIOSUS - A FEASIBILITY STUDY IN DOGS [J].
MAGAL, C ;
WRIGHT, KC ;
DUPRAT, G ;
WALLACE, S ;
GIANTURCO, C .
INVESTIGATIVE RADIOLOGY, 1989, 24 (04) :272-276