4-CHANNEL TELEMETRY SYSTEM FOR INVIVO MEASUREMENT OF HIP-JOINT FORCES

被引:43
作者
GRAICHEN, F
BERGMANN, G
机构
[1] Free University, Department of Orthopaedics, Biomechanics-Lab., D-1000 Berlin 33
来源
JOURNAL OF BIOMEDICAL ENGINEERING | 1991年 / 13卷 / 05期
关键词
IMPLANTABLE TELEMETRY; INSTRUMENTED HIP PROSTHESES; JOINT FORCES; BIOMECHANICS;
D O I
10.1016/0141-5425(91)90016-Z
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The long-term loosening of artificial hip joints remains a serious clinical problem. Optimization of implant design and material will improve the fixation, but it requires a detailed knowledge of the forces which act on the implant. A four-channel telemetric transmitter was developed and arranged completely inside a hermetically closed artificial hip joint. This permits long term in vivo measurements of the three-dimensional forces without endangering the patient. The external telemetry system consists of an inductive power supply, an RF receiver, a microcomputer with hardware extension and a VHS video system. The personal computer offers real-time data processing of three orthogonal force components as well as slow motion analysis of recorded measurements. After several years of animal tests, two instrumented prostheses were implanted in the first patient (male) in May and August of 1988. In March 1990 a third prosthesis was implanted in a second patient (female). Joint force measurements have regularly been performed from the first post-operative day until now for several kinds of activity.
引用
收藏
页码:370 / 375
页数:6
相关论文
共 16 条
[1]   INVIVO MEASUREMENT OF HIP-JOINT LOADING .1. PHYSIOTHERAPY [J].
BERGMANN, G ;
ROHLMANN, A ;
GRAICHEN, F .
ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 1989, 127 (06) :672-679
[2]   MULTICHANNEL STRAIN-GAUGE TELEMETRY FOR ORTHOPEDIC IMPLANTS [J].
BERGMANN, G ;
GRAICHEN, F ;
SIRAKY, J ;
JENDRZYNSKI, H ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1988, 21 (02) :169-176
[3]  
BERGMANN G, 1990, 36TH T ORS NEW ORL, P2
[4]  
BERGMANN G, 1990, IN PRESS J BIOMECH
[5]  
BERGMANN G, 1989, 12 C INT SOC BIOM LO
[6]  
BERGMANN G, 1982, 9TH P WORLD C IMECO, P395
[7]  
BERGMANN G, 1991, 37 SUBM ANN M ORTH R
[8]  
Bergmann G, 1990, IMPLANTABLE TELEMETR, P35
[9]  
BERGMANN G, 1990, 7TH P M EUR SOC BIOM
[10]   TELEMETRIC FORCE MEASUREMENTS ACROSS THE HIP AFTER TOTAL ARTHROPLASTY [J].
DAVY, DT ;
KOTZAR, GM ;
BROWN, RH ;
HEIPLE, KG ;
GOLDBERG, VM ;
HEIPLE, KG ;
BERILLA, J ;
BURSTEIN, AH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (01) :45-50