A PIEZOELECTRIC IMMUNOSENSOR FOR HEPATITIS VIRUSES

被引:52
作者
KONIG, B
GRATZEL, M
机构
[1] Institut de Chimie Physique II, Ecole Polytechnique Fédérale de Lausanne
关键词
PIEZOELECTRIC SENSORS; BIOSENSORS; HEPATITIS VIRUSES; VIRUS SPECIFIC MONOCLONAL ANTIBODIES; ELISA;
D O I
10.1016/0003-2670(94)00640-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A reusable piezoelectric immunosensor was used for the detection of the human hepatitis viruses type A and type B. The virus specific monoclonal antibodies were immobilized via protein A onto a 16-MHz AT-cut crystal. Binding of the virions resulted for 1 x 10(5) to 1 x 10(10) viruses in solution in a linear frequency change. The detection limit of our system is 1 x 10(5) virions and the experimental error is 8%. When the modified crystal was stored dry over silica gel blue at room temperature, the long term stability of our system was four weeks. Under these conditions, the coated crystal can be used ten times without detectable loss of activity. These results are compared with those for an enzyme-linked immunosorbent assay (ELISA).
引用
收藏
页码:19 / 25
页数:7
相关论文
共 20 条
  • [1] ATHERTON E, 1987, PEPTIDES ANAL SYNTHE, V9, P1
  • [2] BERGMEYER HU, 1986, METHOD ENZYMAT AN, P448
  • [3] BONIN O, 1973, QUANTITATIVE VIROLOG, P1
  • [4] BURKHARDT F, 1992, MIKROBIOLOGISCHE DIA, P356
  • [5] CAMPBELL AM, 1991, MONOCLONAL ANTIBODY, P63
  • [6] GINSBERG HS, 1988, VIROLOGY, P321
  • [7] NOVEL IMMUNOSENSORS
    KARUBE, I
    SUZUKI, M
    [J]. BIOSENSORS, 1986, 2 (06) : 343 - 362
  • [8] KONIG B, 1993, ANAL LETT, V26, P2313
  • [9] KONIG B, 1993, ANAL LETT, V26, P1567, DOI 10.1080/00032719308021481
  • [10] LONG-TERM STABILITY AND IMPROVED REUSABILITY OF A PIEZOELECTRIC IMMUNOSENSOR FOR HUMAN ERYTHROCYTES
    KONIG, B
    GRATZEL, M
    [J]. ANALYTICA CHIMICA ACTA, 1993, 280 (01) : 37 - 41