ELIMINATION OF Z-EJECTION IN FOURIER-TRANSFORM ION-CYCLOTRON RESONANCE MASS-SPECTROMETRY BY RADIO-FREQUENCY ELECTRIC-FIELD SHIMMING

被引:60
作者
WANG, MD [1 ]
MARSHALL, AG [1 ]
机构
[1] OHIO STATE UNIV,DEPT CHEM,120 W 18TH AVE,COLUMBUS,OH 43210
关键词
D O I
10.1021/ac00204a017
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In Fourier transform ion cyclotron resonance (FT/ICR) mass spectrometry, coherent Ion cyclotron orbital motion is produced by resonant radio frequency (rf) electric field excitation. However, because the excitation electrodes are of finite dimensions, the desired transverse (to the applied magnetic field) rf electric field Is accompanied by an rf electric field component along the z- (magnetic field) direction, resulting in mass-dependent z -ejection and mass-dependent FT/ICR mass spectral peak relative magnitudes. Addition of several “guard wires” of voltage-divided rf amplitude allows the rf electric field to be “shimmed” to near-perfect uniformity. In this paper (see also the accompanying paper by Russell et al.), we introduce two types of rf-shlmmed ion traps. In the first type, guard wires are placed only in front of the trapping electrodes. In the second type, guard wire rings are placed inside the detector and trapping electrodes. For either arrangement, SIMION simulations were used to adjust the rf voltages applied (by use of voltage dividers) to the guard wires or rings so as to produce an optimally uniform rf field within the trap. The virtual elimination of z-excitation Is confirmed by plots of magnitude-mode relative peak height vs ICR orbital radius. Because the guard wires (or rings) tend to shield the Ions from the trapping electrode potential, the shift In ICR frequency with trapping voltage Is also reduced, but not as well as by a screened trap. In slde-by-side tests of conventional and rf-shlmmed traps mounted collinearly along the z-axls, the ICR signal strength was reduced to ~87% and ~25% of Its unshimmed level by the addition of guard wires or guard rings, respectively. © 1990, American Chemical Society. All rights reserved.
引用
收藏
页码:515 / 520
页数:6
相关论文
共 41 条
[1]   EJECTION OF LOW-MASS CHARGED-PARTICLES IN HIGH MAGNETIC-FIELD ICR SPECTROMETERS [J].
ALLEMANN, M ;
KOFEL, P ;
KELLERHALS, H ;
WANCZEK, KP .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1987, 75 (01) :47-54
[2]  
AMSTER IJ, 1986, ANAL CHEM, V58, P483
[3]  
ASAMOTO B, 1988, SPECTROSCOPY, V3, P38
[4]   AN ION EJECTION TECHNIQUE FOR STUDY OF ION-MOLECULE REACTIONS WITH ION CYCLOTRON RESONANCE SPECTROSCOPY [J].
BEAUCHAMP, JL ;
ARMSTRONG, JT .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1969, 40 (01) :123-+
[5]  
BUCHANAN MV, ACS SYM SER, V359, P1
[6]  
CHEN L, 1986, CHEMOMETRICS INTELLI, V1, P51
[7]  
CODY RB, IN PRESS LASERS MASS
[8]   FREQUENCY-SWEEP FOURIER-TRANSFORM ION-CYCLOTRON RESONANCE SPECTROSCOPY [J].
COMISARO.MB ;
MARSHALL, AG .
CHEMICAL PHYSICS LETTERS, 1974, 26 (04) :489-490
[9]   FOURIER-TRANSFORM ION-CYCLOTRON RESONANCE SPECTROSCOPY [J].
COMISAROW, MB ;
MARSHALL, AG .
CHEMICAL PHYSICS LETTERS, 1974, 25 (02) :282-283
[10]   CUBIC TRAPPED-ION CELL FOR ION-CYCLOTRON RESONANCE [J].
COMISAROW, MB .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1981, 37 (02) :251-257