THE REPRODUCIBILITY OF TEST CONTRACTIONS FOR CALIBRATION OF ELECTROMYOGRAPHIC MEASUREMENTS

被引:72
作者
VEIERSTED, KB
机构
[1] National Institute of Occupational Health, Oslo 1, N-0033
来源
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY AND OCCUPATIONAL PHYSIOLOGY | 1991年 / 62卷 / 02期
关键词
REPRODUCIBILITY; ELECTROMYOGRAPHY; UPPER TRAPEZIUS MUSCLE; TEST CONTRACTION; ELECTRODE POSITION; CALIBRATION;
D O I
10.1007/BF00626762
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
The main purpose of this study was to evaluate the reproducibility of electromyographic (EMG) measurements and specifically to test a calibration procedure with submaximal test contractions. Bipolar surface electrodes (20 mm fixed distance) were repositioned by a tracing sheet on both trapezius muscles, halfway between acromion and processus prominens. Submaximal test contractions were performed by keeping both arms straight abducted 90-degrees and forward flexed 10-degrees for 15-s periods. The arm position could be precisely reproduced in the frontal plane, but deviated forwards by 4-degrees in the horizontal plane, where the sensitivity of the EMG response to arm position was lowest. The electrodes were repositioned within a radius of 3 mm with a probability of 90%. Large deviations in the EMG response were found within this radius and a significant depression of the EMG response was recorded over the middle part of the muscle (the innervation zone?). This change in sensitivity of the EMG response with electrode position occurred in parallel for the test and maximal contractions. The total coefficient of variation was estimated to be 23% for recurrent EMG measurements using the calibration procedure described.
引用
收藏
页码:91 / 98
页数:8
相关论文
共 24 条
[1]  
[Anonymous], MANUAL PSYCHOPHYSIOL
[2]  
ANTII CJ, 1976, BIOMECHANICS VA, P515
[3]   TOPOGRAPHICAL LOCALIZATION OF MOTOR ENDPLATES IN CRYOSECTIONS OF WHOLE HUMAN MUSCLES [J].
AQUILONIUS, SM ;
ASKMARK, H ;
GILLBERG, PG ;
NANDEDKAR, S ;
OLSSON, Y ;
STALBERG, E .
MUSCLE & NERVE, 1984, 7 (04) :287-293
[4]  
Basmajian JV, 1985, MUSCLE ALIVE THEIR F, V5th
[5]   METABOLIC CHANGES FOLLOWING ECCENTRIC EXERCISE IN TRAINED AND UNTRAINED MEN [J].
EVANS, WJ ;
MEREDITH, CN ;
CANNON, JG ;
DINARELLO, CA ;
FRONTERA, WR ;
HUGHES, VA ;
JONES, BH ;
KNUTTGEN, HG .
JOURNAL OF APPLIED PHYSIOLOGY, 1986, 61 (05) :1864-1868
[6]  
HAGBERG M, 1981, AM J PHYS MED REHAB, V60, P111
[7]  
HAGBERG M, 1975, Electromyography and Clinical Neurophysiology, V15, P253
[8]  
HAGBERG M, 1981, ARBETE HALSA, V24, P1
[9]   EXPERIMENTAL HUMAN-MUSCLE DAMAGE - MORPHOLOGICAL-CHANGES IN RELATION TO OTHER INDEXES OF DAMAGE [J].
JONES, DA ;
NEWHAM, DJ ;
ROUND, JM ;
TOLFREE, SEJ .
JOURNAL OF PHYSIOLOGY-LONDON, 1986, 375 :435-448
[10]  
JONSSON B, 1978, CONT CLIN NEUROPHYSI, V34, P417