EFFECTS OF SOME TECHNOLOGICAL ASPECTS ON THE FATIGUE-STRENGTH OF A CEMENTLESS HIP STEM

被引:19
作者
VICECONTI, M
TONI, A
GIUNTI, A
机构
[1] UNIV BOLOGNA,IST ORTOPED RIZZOLI,BIOMAT TECHNOL LAB,BOLOGNA,ITALY
[2] UNIV BOLOGNA,IST ORTOPED RIZZOLI,ORTHOPAED CLIN,BOLOGNA,ITALY
来源
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH | 1995年 / 29卷 / 07期
关键词
D O I
10.1002/jbm.820290713
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Four prototype cementless hip stems were tested following the ISO 7206 protocol for the assessment of the endurance properties and compared with a cast Cr-Co-Mo (ASTM F75) commercially available stem which was used as reference design. All the tested stems were similar in shape and size, but with some substantial differences. The first was made of forged Ti6Al4V alloy (ASTM F136). The second, made of the same material, featured a central hole intended to reduce the bending stiffness of the stem itself. The third was identical to the second but for a small tooling notch in one of the fillets of the hole. The fourth was similar to the first but had a coating of sintered titanium beads in the proximal part. All of these modifications were made to evaluate the effect of fatigue strength of intentional or unintentional features commonly found in commercial stems. The forged Ti6Al4V alloy was found to be substantially stronger than the cast ASTM F75 Cr-Co-Mo alloy. However, tooling notches or sintered coatings were found to dramatically reduce this strength. Thus, the Ti6Al4V alloy calls for an accurate design process, especially when complex shapes or sintered structures are required. (C) 1995 John Wiley and Sons, Inc.
引用
收藏
页码:875 / 881
页数:7
相关论文
共 15 条
[1]  
Black J., 1988, ORTHOPAEDIC BIOMATER
[2]  
CHARNLEY J, 1975, CLIN ORTHOP RELAT R, V111, P105
[3]   FATIGUE PROPERTIES OF IMPLANT MATERIALS IN HIP-PROSTHESIS FORM - A STANDARDIZED TEST [J].
DUCHEYNE, P ;
WEVERS, M ;
DEMEESTER, P .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1983, 17 (01) :45-57
[4]  
ENGH CA, 1989, CLIN ORTHOP RELAT R, P141
[5]   5 YEARS EXPERIENCE WITH CERAMIC-METAL-COMPOSITE HIP ENDOPROSTHESES .1. CLINICAL-EVALUATION [J].
GRISS, P ;
HEIMKE, G .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1981, 98 (03) :157-164
[6]  
HUMPHREYS PK, 1980, P I MECH ENG, V204, P29
[7]   FRACTURE OF HIP PROSTHESES DUE TO WELDING DEFECTS [J].
KOUSBROEK, R ;
DUCHEYNE, P ;
VANDERPERRE, G ;
AERNOUDT, E ;
MULIER, JC .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1981, 98 (01) :51-56
[8]  
MARMOR L, 1984, CLIN ORTHOP RELAT R, P148
[9]   FRACTURE FAILURE OF A FORGED VITALLIUM PROSTHESIS - A CASE-REPORT [J].
MILLER, EH ;
SHASTRI, R ;
SHIH, CI .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1982, 64 (09) :1359-1363
[10]  
MUNUERA L, 1992, CLIN ORTHOP RELAT R, P163