CHARACTERIZATION AND EVALUATION OF A NEW DOUBLE LUMEN CENTRAL VENOUS CATHETER

被引:4
作者
DONNELLY, PK
MORTON, D
WOFFINDIN, C
HOENICH, NA
机构
[1] UNIV LEICESTER,DEPT SURG,LEICESTER LE1 7RH,ENGLAND
[2] UNIV NEWCASTLE UPON TYNE,SCH CLIN MED SCI,DEPT MED,NEWCASTLE TYNE NE1 7RU,TYNE & WEAR,ENGLAND
关键词
NEW CENTRAL VENOUS CATHETER; HEMODIALYSIS; LABORATORY STUDIES;
D O I
10.1177/039139889201500603
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The performance of a new double lumen central venous haemodialysis catheter was tested in two laboratory models. In a bench model the patient's venous system was simulated by a reservoir from which water or glycerol was drawn through a fixed tube. A double lumen silastic catheter was then inserted into the tube, as it would in a major vein, with the tip directed away from the direction of flow. The catheter was linked to a dialysis circuit incorporating pressure sensors and dye was infused at constant rate so that recirculation at the tip could be measured and found to be less than 5%. The same catheters were inserted operatively into the superior vena cava via the external jugular vein of three pigs (weight 27-31 kg). The catheters remained patent for four weeks and when connected to an extracorporeal circuit had recirculation and pressure flow characteristics comparable to the bench model in the range 50-400 ml/min. The new double lumen catheter is worthy of clinical evaluation.
引用
收藏
页码:331 / 338
页数:8
相关论文
共 26 条
[1]  
BREGMAN H, 1986, T AM SOC ART INT ORG, V32, P500
[2]   A SURVEY OF VASCULAR ACCESS FOR HEMODIALYSIS IN THE NETHERLANDS [J].
BURGER, H ;
KOOTSTRA, G ;
DECHARRO, F ;
LEFFERS, P .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 1991, 6 (01) :5-10
[3]   INTERNAL JUGULAR-VEIN CANNULATION USING 2 SILASTIC CATHETERS - A NEW, SIMPLE AND SAFE LONG-TERM VASCULAR ACCESS FOR EXTRACORPOREAL TREATMENT [J].
CANAUD, B ;
BERAUD, JJ ;
JOYEUX, H ;
MION, C .
NEPHRON, 1986, 43 (02) :133-138
[4]   CENTRAL VENOUS ACCESS FOR HEMODIALYSIS USING THE HICKMAN CATHETER [J].
CAPPELLO, M ;
DEPAUW, L ;
BASTIN, G ;
PROSPERT, F ;
DELCOUR, C ;
THAYSSE, C ;
DHAENE, M ;
VANHERWEGHEM, JL ;
KINNAERT, P .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 1989, 4 (11) :988-992
[5]  
CHARM SE, 1974, BLOOD FLOW MICROCIRU
[6]   SUPERIORITY OF THE INTERNAL JUGULAR OVER THE SUBCLAVIAN ACCESS FOR TEMPORARY DIALYSIS [J].
CIMOCHOWSKI, GE ;
WORLEY, E ;
RUTHERFORD, WE ;
SARTAIN, J ;
BLONDIN, J ;
HARTER, H .
NEPHRON, 1990, 54 (02) :154-161
[7]  
DEMERS HG, 1989, DIALYSIS TRANSPLANT, V18, P4130
[8]  
Donnelly P K, 1988, Nephrol Dial Transplant, V3, P57
[9]  
DONNELLY PK, 1990, INTENSIVE THERAPY CL, P214
[10]  
DUNN J, 1987, SURGERY, V102, P784