CLINICAL-PERFORMANCE OF THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR - ELECTROCARDIOGRAPHIC DOCUMENTATION OF 101 SPONTANEOUS DISCHARGES

被引:57
作者
MALONEY, J
MASTERSON, M
KHOURY, D
TROHMAN, R
WILKOFF, B
SIMMONS, T
MORANT, V
CASTLE, L
机构
[1] Department of Cardiology, Cleveland, Ohio
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1991年 / 14卷 / 02期
关键词
HOLTER MONITORING; IMPLANTABLE DEFIBRILLATOR PERFORMANCE; SPONTANEOUS DISCHARGES;
D O I
10.1111/j.1540-8159.1991.tb05107.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Records of 105 patients, who received an automatic implantable cardioverter defibrillator (AICD), were studied to investigate the causes of spontaneous AICD discharges and to correlate the symptoms with the arrhythmias triggering AICD discharges. During a follow-up period of 13 +/- 8 months, 46/105 (44%) patients had 566 spontaneous AICD discharges. A total of 101 discharges were documented with Holter monitoring in 23 patients. In this study group, there were 8 (8%) AICD discharges for 5 episodes of ventricular fibrillation, and 68 (67%) discharges for 63 episodes of sustained ventricular tachycardia. Patients lost consciousness in all episodes of ventricular fibrillation, but were symptomatic prior to only 36 (53%) discharges in ventricular tachycardia. Non-sustained ventricular tachycardia persisting for a period of 7.5 +/- 2 seconds resulted in 20 AICD discharges; patients were symptomatic prior to 13 (65%) discharges. Supraventricular tachycardias triggered three discharges. One patient had two spurious discharges during sinus rhythm. In conclusion, most of the spontaneous AICD discharges were appropriate for the detected rhythms, but only clinically appropriate for the management of arrhythmias in 75% of the cases. A significant portion of the patients with sustained or nonsustained ventricular tachycardias triggering AICD discharges were asymptomatic prior to discharge, which requires further assessment of the physiology of the arrhythmia as a component of the detection algorithm.
引用
收藏
页码:280 / 285
页数:6
相关论文
共 14 条
[1]   THE USE OF BEEP-O-GRAMS IN THE ASSESSMENT OF AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR SENSING FUNCTION [J].
BALLAS, SL ;
RASHIDI, R ;
MCALISTER, H ;
CORBELLI, R ;
MCCOWAN, R ;
WILKOFF, BL ;
CASTLE, LW ;
MORANT, VA ;
SIMMONS, TW ;
MALONEY, JD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (11) :1737-1745
[2]   THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - CLINICAL-EXPERIENCE, COMPLICATIONS, AND FOLLOW-UP IN 25 PATIENTS [J].
BORBOLA, J ;
DENES, P ;
EZRI, MD ;
HAUSER, RG ;
SERRY, C ;
GOLDIN, MD .
ARCHIVES OF INTERNAL MEDICINE, 1988, 148 (01) :70-76
[3]   CLINICAL-EXPERIENCE, COMPLICATIONS, AND SURVIVAL IN 70 PATIENTS WITH THE AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR [J].
ECHT, DS ;
ARMSTRONG, K ;
SCHMIDT, P ;
OYER, PE ;
STINSON, EB ;
WINKLE, RA .
CIRCULATION, 1985, 71 (02) :289-296
[4]   CONTINUOUS RIGHT VENTRICULAR VOLUME ASSESSMENT BY CATHETER MEASUREMENT OF IMPEDANCE FOR ANTITACHYCARDIA SYSTEM CONTROL [J].
KHOURY, D ;
MCALISTER, H ;
WILKOFF, B ;
SIMMONS, T ;
RUDY, Y ;
MCCOWAN, R ;
MORANT, V ;
CASTLE, L ;
MALONEY, J .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (12) :1918-1926
[5]   AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR - INADVERTENT DISCHARGES DURING PERMANENT PACEMAKER MAGNET TESTS [J].
KIM, SG ;
FURMAN, S ;
MATOS, JA ;
WASPE, LE ;
BRODMAN, R ;
FISHER, JD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1987, 10 (03) :579-582
[6]   ERRONEOUS DISCHARGE FROM AN IMPLANTED AUTOMATIC DEFIBRILLATOR DURING SUPRAVENTRICULAR TACHYARRHYTHMIA INDUCED VENTRICULAR-FIBRILLATION [J].
MANZ, M ;
GERCKENS, U ;
LUDERITZ, B .
AMERICAN JOURNAL OF CARDIOLOGY, 1986, 57 (04) :343-344
[7]   THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - EFFICACY, COMPLICATIONS, AND DEVICE FAILURES [J].
MARCHLINSKI, FE ;
FLORES, BT ;
BUXTON, AE ;
HARGROVE, WC ;
ADDONIZIO, VP ;
STEPHENSON, LW ;
HARKEN, AH ;
DOHERTY, JU ;
GROGAN, EW ;
JOSEPHSON, ME .
ANNALS OF INTERNAL MEDICINE, 1986, 104 (04) :481-488
[8]  
MINER PG, 1985, AM J CARDIOL, V56, P588
[9]   THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - AN OVERVIEW [J].
MIROWSKI, M .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1985, 6 (02) :461-466
[10]   SUDDEN-DEATH RECORDED DURING HOLTER MONITORING [J].
NIKOLIC, G ;
BISHOP, RL ;
SINGH, JB .
CIRCULATION, 1982, 66 (01) :218-225