BETA-PARTICLE DOSIMETRY IN RADIATION SYNOVECTOMY

被引:83
作者
JOHNSON, LS
YANCH, JC
SHORTKROFF, S
BARNES, CL
SPITZER, AI
SLEDGE, CB
机构
[1] MIT,DEPT NUCL ENGN,CAMBRIDGE,MA 02139
[2] MIT,WHITAKER COLL HLTH SCI & TECHNOL,CAMBRIDGE,MA 02139
[3] BRIGHAM & WOMENS HOSP,DEPT ORTHOPED SURG,BOSTON,MA 02115
来源
EUROPEAN JOURNAL OF NUCLEAR MEDICINE | 1995年 / 22卷 / 09期
关键词
DOSIMETRY; RADIATION SYNOVECTOMY; MONTE CARLO SIMULATION; RHEUMATOID ARTHRITIS;
D O I
10.1007/BF00808408
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Beta-particle dosimetry of various radionuclides used in the treatment of rheumatoid arthritis was estimated using Monte Carlo radiation transport simulation coupled with experiments using reactor-produced radionuclides and radiachromic film dosimeters inserted into joint phantoms and the knees of cadavers. Results are presented as absorbed dose factors (cGy-cm(2)/MBq-s) versus depth in a mathematical model of the rheumatoid joint which includes regions of bone, articular cartilage, joint capsule, and tissue (synovium) found in all synovial joints. The factors can be used to estimate absorbed dose and dose rate distributions in treated joints. In particular, guidance is provided for those interested in (a) a given radionuclide's therapeutic range, (b) the amount of radioactivity to administer on a case-by-case basis, (c) the expected therapeutic dose to synovium, and (d) the radiation dose imparted to other, nontarget components in the joint, including bone and articular cartilage.
引用
收藏
页码:977 / 988
页数:12
相关论文
共 35 条
[1]   EVALUATION OF INTRA-ARTICULAR COLLOIDAL GOLD AU 198 IN TREATMENT OF PERSISTENT KNEE EFFUSIONS [J].
ANSELL, BM ;
CROOK, A ;
BYWATERS, EG ;
MALLARD, JR .
ANNALS OF THE RHEUMATIC DISEASES, 1963, 22 (06) :435-&
[2]   YTTRIUM-90 CITRATE COLLOID FOR RADIOISOTOPE SYNOVECTOMY [J].
BOWEN, BM ;
DARRACOTT, J ;
GARNETT, ES ;
TOMLINSON, RH .
AMERICAN JOURNAL OF HOSPITAL PHARMACY, 1975, 32 (10) :1027-1030
[3]  
BRODACK J W, 1992, Journal of Nuclear Medicine, V33, P980
[4]   Phototropy [J].
Chalkley, L .
CHEMICAL REVIEWS, 1929, 6 (02) :217-280
[5]   A SOLID WATER PHANTOM MATERIAL FOR RADIOTHERAPY X-RAY AND GAMMA-RAY BEAM CALIBRATIONS [J].
CONSTANTINOU, C ;
ATTIX, FH ;
PALIWAL, BR .
MEDICAL PHYSICS, 1982, 9 (03) :436-441
[6]  
Constantinou C, 1978, THESIS U LONDON
[7]  
DECKART H, 1979, RADIOBIOL RADIOTHER, V3, P363
[8]  
DEUTSCH E, 1993, EUR J NUCL MED, V20, P1113
[9]  
FELLINGER K, 1952, Wien Z Inn Med, V33, P351
[10]   USE OF YTTRIUM 90 IN PERSISTENT SYNOVITIS OF KNEE .1. RETENTION IN KNEE AND SPREAD IN BODY AFTER INJECTION [J].
GUMPEL, JM ;
WILLIAMS, ED ;
GLASS, HI .
ANNALS OF THE RHEUMATIC DISEASES, 1973, 32 (03) :223-227