PORTAL DOSIMETRY USING A LIQUID ION-CHAMBER MATRIX - DOSE-RESPONSE STUDIES

被引:56
作者
ZHU, YP
JIANG, XQ
VANDYK, J
机构
[1] Department of Clinical Physics, Ontario Cancer Institute, Toronto, Ontario
关键词
X RAY; CALIBRATION; ION CHAMBER MATRIX; PORTAL DOSIMETRY; DOSE VERIFICATION;
D O I
10.1118/1.597502
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Current intensive investigations of electronic portal imaging devices (EPIDs) have prompted their potential application to portal dosimetry. In this paper, the progress made in using a commercial liquid ion chamber matrix EPID for portal dosimetry is discussed. The pixel value of the liquid ion chamber element was calibrated against dose by exposing the imager to 6-MV x-ray beams of various intensities obtained with various thicknesses of lead attenuators and a range of source to detector distances. Absolute dose values were determined using an ion chamber on the central axis at the depth of maximum dose in a solid water phantom. The pixel values of the matrix were determined for various field sizes in order to evaluate the dependence of pixel value on dose at those field sizes. It was confirmed that the pixel value was proportional to the square root of the dose rate and was nearly independent of the field size. The 2D pixel values were converted to 2D dose maps in the water phantom after applying a correction for the effect of horns in the flood calibration field. The flood calibration field was used to obtain the relative sensitivity of each pixel. Good agreement was observed (normally better than 1% in relative standard deviation) between the converted dose distribution obtained from the pixel matrix and the direct dose measurement using an ion chamber scanned in a water phantom in regions of shallow dose gradient. For application to on-line portal dosimetry, both the short- and long-term stability of this EPID system were found to be within 1% relative standard deviation. This system, together with an accurate portal dose calculation procedure, can be used for on-line radiotherapy dose verification.
引用
收藏
页码:1101 / 1106
页数:6
相关论文
共 21 条
  • [1] POTENTIAL AND LIMITATIONS OF INVARIANT KERNEL CONFORMAL THERAPY
    BOYER, AL
    DESOBRY, GE
    WELLS, NH
    [J]. MEDICAL PHYSICS, 1991, 18 (04) : 703 - 712
  • [2] A REVIEW OF ELECTRONIC PORTAL IMAGING DEVICES (EPIDS)
    BOYER, AL
    ANTONUK, L
    FENSTER, A
    VANHERK, M
    MEERTENS, H
    MUNRO, P
    REINSTEIN, LE
    WONG, J
    [J]. MEDICAL PHYSICS, 1992, 19 (01) : 1 - 16
  • [3] OPTIMIZATION OF STATIONARY AND MOVING BEAM RADIATION-THERAPY TECHNIQUES
    BRAHME, A
    [J]. RADIOTHERAPY AND ONCOLOGY, 1988, 12 (02) : 129 - 140
  • [4] DOSE CALCULATION FOR PHOTON BEAMS WITH INTENSITY MODULATION GENERATED BY DYNAMIC JAW OR MULTILEAF COLLIMATIONS
    CHUI, CS
    LOSASSO, T
    SPIROU, S
    [J]. MEDICAL PHYSICS, 1994, 21 (08) : 1237 - 1244
  • [5] EVANS PM, 11TH P INT C COMP RA, P118
  • [6] TRANSMISSION DOSE MEASUREMENT FOR COBALT-60 RADIATION WITH SPECIAL REFERENCE TO ROTATION THERAPY
    FEDORUK, SO
    JOHNS, HE
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1957, 30 (352) : 190 - 195
  • [7] EXIT DOSE MEASUREMENTS BY PORTAL FILM DOSIMETRY
    FIORINO, C
    DELVECCHIO, A
    CATTANEO, GM
    FUSCA, M
    LONGOBARDI, B
    SIGNOROTTO, P
    CALANDRINO, R
    [J]. RADIOTHERAPY AND ONCOLOGY, 1993, 29 (03) : 336 - 340
  • [8] GALVIN JM, 1992, INT J RADIAT ONCOL, V27, P789
  • [9] HANSEN VN, UNPUB 11TH P INT C C, P116
  • [10] HEIJMEN BJ, 1994, 11TH P INT C COMP RA, P112