DIGITAL AUDIO TAPE AS A METHOD OF STORING DOPPLER ULTRASOUND SIGNALS

被引:9
作者
BUSH, G
EVANS, DH
机构
[1] Sch. of Med., Leicester Univ.
关键词
D O I
10.1088/0967-3334/14/3/017
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Digital audio tape recorders perform significantly better than conventional tape recorders in respect of their frequency response, dynamic range, signal-to-noise ratio, and wow and flutter figures. They also have advantages in terms of their tape size and playing time, and their ability to record additional information as 'sub-codes'. The suitability of DAT recorders as a means for storing ultrasound Doppler signals has been evaluated. The results show that they are an ideal instrument for recording directional Doppler signals, but like conventional tape recorders are not suitable for storing quadrature Doppler signals.
引用
收藏
页码:381 / 386
页数:6
相关论文
共 3 条
[1]   RECORDING DOPPLER BLOOD-FLOW SIGNALS ON MAGNETIC-TAPE [J].
SMALLWOOD, RH .
CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1985, 6 (04) :357-359
[2]   SIMPLE SYSTEM FOR AUTOMATIC INTERMITTENT RECORDING OF BLOOD-FLOW IN FEMORODISTAL BYPASS GRAFTS USING DOPPLER ULTRASOUND [J].
THRUSH, AJ ;
EVANS, DH .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1990, 28 (02) :193-195
[3]  
VANS DH, 1989, DOPPLER ULTRASOUND P, pCH6