SURFACE COATING OF HYDROPHILIC-HYDROPHOBIC BLOCK-COPOLYMERS ON A POLY(ACRYLONITRILE) HEMODIALYZER REDUCES PLATELET-ADHESION AND ITS TRANSMEMBRANE STIMULATION

被引:36
作者
MATSUDA, T [1 ]
ITO, S [1 ]
机构
[1] OSAKA CITY UNIV,SCH MED,DEPT UROL,OSAKA 563,JAPAN
关键词
HEMODIALYSIS; MEMBRANES; POLY(ACRYLONITRILE); BIOCOMPATIBILITY;
D O I
10.1016/0142-9612(94)90219-4
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Surface design aimed at reduced adhesion and preserved functions of platelets is of great importance for extracorporeal devices. In this study, a coating technique using hydrophilic-hydrophobic block co-polymers on a hydrophobic poly(acrylonitrile) (PAN) haemodialyser was explored. The hydrophilic block of co-polymers was composed of either poly(methoxy polyethylene glycol methacrylate) or poly(dimethyl acrylamide), and the hydrophobic block was poly(methyl methacrylate). The co-polymers were coated on the dialyser membrane by means of a solution coating method. Upon coating, the hydrophobic block of the co-polymers was anchored on a PAN membrane and the hydrophilic block oriented towards the blood-material interface. This was deduced from water wettability measurements. Significantly reduced transmembrane stimulation of platelets was observed, which was evaluated by determining the intracellular calcium ion concentration of platelets eluted through treated hollow fibres. This suppression was enhanced as the relative fraction of the hydrophilic block of the co-polymers increased. Furthermore, the number of platelets adhering to the co-polymer-coated PAN membrane was drastically reduced. Thus, coating of the hydrophilic-hydrophilic block co-polymers provided better biocompatibility on a hydrophobic PAN dialyser.
引用
收藏
页码:417 / 422
页数:6
相关论文
共 18 条
[1]  
Akizawa T, 1989, ASAIO Trans, V35, P333
[2]  
AKIZAWA T, 1986, T AM SOC ART INT ORG, V32, P76
[3]  
ANDRADE JD, 1985, SURFACE INTERFACIAL, V1, P214
[4]   COMPLEMENT AND LEUKOCYTE-MEDIATED PULMONARY DYSFUNCTION IN HEMODIALYSIS [J].
CRADDOCK, PR ;
FEHR, J ;
BRIGHAM, KL ;
KRONENBERG, RS ;
JACOB, HS .
NEW ENGLAND JOURNAL OF MEDICINE, 1977, 296 (14) :769-774
[5]   COMPLEMENT (C5A)-INDUCED GRANULOCYTE AGGREGATION INVITRO - POSSIBLE MECHANISM OF COMPLEMENT-MEDIATED LEUKOSTASIS AND LEUKOPENIA [J].
CRADDOCK, PR ;
HAMMERSCHMIDT, D ;
WHITE, JG ;
DALMASSO, AP ;
JACOB, HS .
JOURNAL OF CLINICAL INVESTIGATION, 1977, 60 (01) :260-264
[6]   HEMODIALYSIS-ASSOCIATED PLATELET LOSS - STUDY OF THE RELATIVE CONTRIBUTION OF DIALYZER MEMBRANE-COMPOSITION AND GEOMETRY [J].
DOCCI, D ;
TURCI, F ;
DELVECCHIO, C ;
BILANCIONI, R ;
CENCIOTTI, L ;
PRETOLANI, E .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1984, 7 (06) :337-340
[7]  
GRYNKIEWICZ G, 1985, J BIOL CHEM, V260, P3440
[8]   A MODIFICATION OF CELLULOSE THAT FACILITATES THE CONTROL OF COMPLEMENT ACTIVATION [J].
JOHNSON, RJ ;
LELAH, MD ;
SUTLIFF, TM ;
BOGGS, DR .
BLOOD PURIFICATION, 1990, 8 (06) :318-328
[9]  
KANBAYASHI J, 1986, JPN J CLIN PATHOL, V71, P118
[10]   INTERACTIONS OF POLY(ETHYLENE GLYCOL)-GRAFTED CELLULOSE MEMBRANES WITH PROTEINS AND PLATELETS [J].
KISHIDA, A ;
MISHIMA, K ;
CORRETGE, E ;
KONISHI, H ;
IKADA, Y .
BIOMATERIALS, 1992, 13 (02) :113-118