A NEW EFFICIENT NANOTIP LEAD

被引:29
作者
STOKES, K
BIRD, T
机构
[1] Medtronic, Inc., Minneapolis, Minnesota
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1990年 / 13卷 / 12期
关键词
STERIOD-ELUTING LEAD; POROUS TIP LEAD;
D O I
10.1111/j.1540-8159.1990.tb06913.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The ideal lead has low, stable acute and chronic thresholds, high pacing impedance, and good sensing. Leads with low, stable thresholds have been developed, but pacing impedance has been in the 600 OMEGA region. One way to increase pacing impedance is to decrease the electrode's surface area. The threshold performance and sensing ability of < 5 mm2 electrodes have been considered questionable, up to now. We have developed a 1.5 mm2 porous, platinized, steroid-eluting electrode and have demonstrated in canine studies that it has excellent performance. Chronic thresholds are low at about 0.65 +/- 0.28 V (ventricular) and 0.42 +/- 0.12 V (atrial) at 0.5 msec. Chronic pacing impedance is in the 1200-1300 OMEGA region, but mean chronic R and P wave source impedance is less-than-or-equal-to 1500 OMEGA. Sensing is excellent, with almost double the P wave amplitudes usually measured in the canine.
引用
收藏
页码:1901 / 1905
页数:5
相关论文
共 13 条
[1]  
ANDERSON JM, 1988, ASAIO, V11, P101
[2]  
BORNZIN GA, 1985, 3 DECADE CARDIAC PAC, P79
[3]   DECREASING ELECTRODE SIZE AND INCREASING EFFICIENCY OF CARDIAC STIMULATION [J].
FURMAN, S ;
PARKER, B ;
ESCHER, DJW .
JOURNAL OF SURGICAL RESEARCH, 1971, 11 (03) :105-&
[4]  
GORDON PL, 1988, PACE, V11, P869
[5]  
HENSON PM, 1980, AM J PATHOL, V101, P494
[6]  
Hoff P.I., 1985, CARDIAC PACING ELECT, P1014
[7]   FAILURE OF DEMAND PACING WITH SMALL SURFACE-AREA ELECTRODES [J].
HUGHES, HC ;
BROWNLEE, RR ;
TYERS, GFO .
CIRCULATION, 1976, 54 (01) :128-132
[8]  
Irnich W., 1973, CARDIAC PACING, P268
[9]   ACUTE VOLTAGE, CHARGE, AND ENERGY THRESHOLDS AS FUNCTIONS OF ELECTRODE SIZE FOR ELECTRICAL-STIMULATION OF THE CANINE HEART [J].
LINDEMANS, FW ;
ZIMMERMAN, ANE .
CARDIOVASCULAR RESEARCH, 1979, 13 (07) :383-391
[10]   THE POROUS TITANIUM STEROID ELUTING ELECTRODE - A DOUBLE-BLIND STUDY ASSESSING THE STIMULATION THRESHOLD EFFECTS OF STEROID [J].
MOND, H ;
STOKES, K ;
HELLAND, J ;
GRIGG, L ;
KERTES, P ;
PATE, B ;
HUNT, D .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (02) :214-219